background image

 

 

Thermo-shaker for microtubes 
and PCR plates 

PCMT Plus 

Operating Manual 

 

 

Содержание PCMT Plus

Страница 1: ...Thermo shaker for microtubes and PCR plates PCMT Plus Operating Manual ...

Страница 2: ...Edition 1 01 December of 2021 Thermo Shaker for Microtubes and PCR Plates PCMT Plus Page 2 Operating manual ...

Страница 3: ... precautions 4 3 General information 5 4 Getting started 6 5 Operation 8 6 Calibration 10 7 Specifications 11 8 Ordering information 12 9 Guarantee and Service 13 10 Compliance 14 1 About this edition of user manual 1 1 The current edition of the user manual applies to the following models Model Version PCMT Plus thermo shaker for microtubes and PCR plates V 1GW ...

Страница 4: ...on the serial num ber label Use only the external power supply provided with this product Ensure that the power switch and external power supply are easily accessible during use Do not plug the unit into an ungrounded power socket and do not use an ungrounded extension lead Disconnect the unit from electric circuit before moving If liquid penetrates into the unit disconnect it from the external po...

Страница 5: ... the user can calibrate the unit approxi mately 6 of the selected temperature to compensate differences in the thermal behaviour of tubes from different manufacturers LCD display indicates pre set and current values of temperature speed and time of operation Quiet motor operation compact size prolonged service life Sensor error handling and diagnostics Bluetooth connectivity control and profiling ...

Страница 6: ...units transported in the original package 4 2 Complete sets 4 2 1 Standard set PCMT Plus thermo shaker 1 pce External power supply 1 pce Power cable 1 pce Spare rubber belt 2 pcs Bluetooth adapter for PC 1 pce Operating manual 1 copy Control software and software manual 1 copy 4 2 2 Optional accessories HC18C thermoblock for microtubes on request HC32 thermoblock for microtubes on request HC24 the...

Страница 7: ...al power supply is disconnect from the device Choose the thermoblock connect the plug to the contact terminal according to the scheme on fig 1 1 on the underside of the thermoblock Make sure that the connector is mounted tightly Align the thermoblock so that the warning label is facing the front of the unit fig 2 Secure with the four knurled screws fig 2 1 or four hex screws 4 5 Changing blocks Di...

Страница 8: ...he 14 digit serial ID number of the unit in the bottom line 5 3 If a temperature is set then the platform temperature that automatically rises to that temperature The time of temperature stabilization depends on the room temperature If the heating of is turned off by setting the temperature below 4 C top line shows indication OFF 5 4 Working with PC Install the software connect and operate the uni...

Страница 9: ... or is reset and the timer indicator in the upper line shows 00 00 pressing the RUN STOP key will start continuous operation of the device with countdown timer in the lower line Actual until the RUN STOP key is pressed again 5 8 The platform motion can be stopped at any time by pressing the RUN STOP key In this case the program realization and the platform motion will stop and the timer will switc...

Страница 10: ...erature in operation mode e g 40 C 6 4 3 After the unit reaches the set temperature when the set and current temperature readings equal leave the unit for 30 min for thermal stabilization 6 4 4 Let us assume that the readings of independent sensor is 39 C but the display s actual temperature is 40 C Then it is necessary to add 1 C correction 6 4 5 Hold the RUN STOP key pressed for more than 8 s to...

Страница 11: ...C 0 1 C Maintaining accuracy2 at 37 C 0 5 C Uniformity over the platform2 at 4 C 0 6 C at 37 C 0 1 C at 100 C 0 3 C Average heating speed from 25 С to 100 С 5 С min Average cooling speed from 100 С to 25 C 5 С min from 25 C to 4 С 1 8 С min Calibration option yes Calibration coefficient range 0 936 1 063 0 063 7 2 General parameters Speed range 250 1400 rpm Speed setting resolution 10 rpm Maximal ...

Страница 12: ...rts contact Grant or your local Grant representative 8 2 1 Optional accessories Model Description Weight1 kg HC18 For 20x0 5 ml 12x1 5ml tubes 0 7 HC32 For 20x0 2 ml 12x1 5ml tubes 0 7 HC24 For 24x2 0 ml microtubes 0 6 HC24N For 24x1 5 ml microtubes 0 7 HC96 For 96 well microplate for PCR w o skirt with half skirt low and high profile 0 7 8 2 2 Replacement parts Replacement part Description Rubber...

Страница 13: ...g solution e g Grant PDS 250 After disinfecting wipe the surfaces dry 9 3 3 The unit and its accessories are not autoclavable 9 4 Rubber belt replacement For maintenance of reliable operation of the device the manufacturer recommends replacing rubber belts after 1 5 years or 2000 hours of operation time Disconnect the external power supply from the device Remove 4 fixation screws on the device bot...

Страница 14: ...pment Regulations 2012 BS EN50581 Waste Electrical and Electronic Equipment WEEE All the products covered by this Manual are marked with the crossed out wheelie bin sym bol indicating they must not be disposed of with unsorted waste Safe recycling of WEEE helps conserve natural resources and protect human health Grant Instruments complies fully with the UK Waste Electrical Electronic Equipment WEE...

Страница 15: ...Thermo Shaker for Microtubes and PCR Plates PCMT Plus Edition 1 01 December of 2021 Operating manual Page 15 ...

Страница 16: ...l salesdesk grantinstruments com www grantinstruments com Representative in the European Union Grant Instruments Europe B V Strawinskylaan 411 WTC Tower A 4th Floor 1077 XX Amsterdam The Netherlands Email grant eu grantinstruments com Thermo Shaker for Microtubes and PCR Plates PCMT Plus 1 01 ...

Отзывы: