User's Guide to the

 

FHP3P01

PON  Power Meter

English Version

FHP3P01

Содержание FHP3P01

Страница 1: ...User s Guide to the FHP3P01 PON Power Meter English Version FHP3P01...

Страница 2: ...User s Guide to the FHP3P01 PON Power Meter...

Страница 3: ...ion 5 Operation 5 1 Layout 5 2 Key s Functions 5 3 Definitions 5 4 Quick Operation 5 5 Detail Operation 5 5 1 Powering On the Instrument 5 5 2 Test mode switch 5 5 3 Threshold Setup 5 5 4 Time Setup 5...

Страница 4: ...d video so called triple play applications along a single fiber FHP3P01 can measure not only 1490nm and 1550nm optical signal but also accurately detect and measure the upstream burst at 1310nm sent f...

Страница 5: ...s of all three wavelengths 1310 1490 1550nm of PON signals with two optical ports Detect and measure the upstream burst at 1310nm Support APON BPON EPON and GPON networks Pass Fail and Warning indicat...

Страница 6: ...on condensing 1 2V 4pcs Ni MH AA 12V AC DC Adapter 3 2v 2pcs lithium battery 2pcs fake battery 20 hours 190mm 105mm 55mm L W H 700g 40dBm 10dBm 50dBm 15dBm 50dBm 20dBm 1310nm 1490nm 1550nm 1260nm 1360...

Страница 7: ...returned to Grandway authorized service center for repair replacement and calibration Exclusions The warranty on your equipment shall not apply to defects resulting from the following Unauthorized re...

Страница 8: ...mer Service 6F Xin an building No 99 Tianzhou Road Shanghai 200233 P R China Contact Us Tel 0086 21 54451260 61 62 63 Fax 0086 21 54451266 E mail overseas grandway com cn Website www grandway com cn F...

Страница 9: ...als and both plugs are fully seated Incomplete engagement may cause fuming electric shock or equipment damage and may result in personal injury Do not operate the equipment near hot objects in hot env...

Страница 10: ...charge batteries 2 Please make sure that you have turned the instrument on before charge the batteries unplug the AC adapter when the batteries are fully charged 3 Please make sure the batteries are w...

Страница 11: ...ent casing into which the output cable of the AC adapter is plugged And when the AC adapter is plugged in the AC Indicator on the LCD will be displayed Note 1 Power is supplied by the AC adapter even...

Страница 12: ...User s Guide Quality Check Report 4 Ni MH Batteries or 4 lithium battery 2 External charger for lithium battery of them is fake battery Carrying Case CD software Optional accessories AC Adapter Checki...

Страница 13: ...ideo OLT ONU Part 2 The front panel is divided to two parts Part I LCD Display Part II Key Matrix The tester shows test result on LCD screen and in the meanwhile indicates different status in part 2 b...

Страница 14: ...is used to set threshold Pressing it down for more than 2 seconds enters threshold setup menu In threshold setup menu press and release the key quickly less than 2 seconds to switch wavelength In time...

Страница 15: ...ring upper limit as Limit1 and lower limit as Limit2 the Pass threshold value in instrument s internal setup are Threshold1 Threshold2 as Warning threshold and Threshold3 as Fail threshold also Limit2...

Страница 16: ...or more than 2 seconds enters to time setup menu Using and REF SEL keys to setup time After the setup press Power key less than 2 sec to save the setting and exit 5 Pressing Threshold key for more tha...

Страница 17: ...In test menu press Power key and release quickly less than 2 seconds to activate or deactivate Auto shut down function The auto shut down function means the instrument will shut down automatically if...

Страница 18: ...er Unit dB is to display a power value relative to reference value In this mode reference value needs to be preset correctly 2 P F test mode P F test mode means the measured light power comparing to p...

Страница 19: ...reshold symbol threshold number and name In this menu press and release Threshold key quickly less than 2 seconds to switch the wavelength among 1310nm 1490nm and 1550nm Press Mode key to view previou...

Страница 20: ...When cursor moves to a number the user can use Mode key to increase value and dBm dB to decrease value When cursor moves to Yes and press REF SEL key for more than 2 seconds the Yes will flash means...

Страница 21: ...ween dBm and dB Here the unit dBm is the actual power and unit dB is a power value relative to reference value Refer to next section about Reference value setup When the instrument is in the P F test...

Страница 22: ...r as reference value The test results afterward are the values after comparing to this reference value Now the unit will be in dB and LED below LCD display will not light In the test menu press and re...

Страница 23: ...hten the backlit and dBm dB key to weaken the backlit After completing the setup press Power key less than 2 sec to save the setting exit and back to the test menu Pressing Threshold and REF SEL keys...

Страница 24: ...xx means the current recording number the maximum number up to 1000 pieces As you may refer to the picture 8 it means there are 24 pieces of test recording inside Press the mode button for longer than...

Страница 25: ...utton Simultaneously press Mode and Threshold to enter the record view interface press the shortly to exit it refer to picture 10 for this interface At the record view interface you may see the main m...

Страница 26: ...ay damage the chip of tester At the record view interface press mode button to move X upward one step press dBm dB button to move X downward one step In picture 10 the interfaces explains that there a...

Страница 27: ...ws full means the battery is fully charged remove the charger 3 Before charging the instrument please make sure the rechargeable battery pack is installed 4 If do not use the instrument for long time...

Страница 28: ...n 2 Make sure that the operating voltage of the AC adapter charger is the same as the local AC line voltage 3 Please always stay the optical connectors away from oil dirt and other contamination to en...

Страница 29: ...26 Interface Threshold setting Tag Calibration setting Tag Power meter setting Tag Upload the data from Confirm the value Press here to save the chart as txt format Press here to s ave the chart as e...

Страница 30: ...eter connect to PC by USB patchcord 2 set standard laser source value 10dBm for 1490nm channel connect laser source with OLT Video port check the certain error 3 Open the software showed in fig1 displ...

Страница 31: ...sh the calibration button wait for seconds display the window in fig4 5 After calibration please check whether value on FHP3P01 has been changed the same procedures as the other wavelengths Caution do...