Gram BIOBASIC 210 Скачать руководство пользователя страница 1

BIOBASIC

USER MANUAL

Models: 210, 310 & 410

Содержание BIOBASIC 210

Страница 1: ...BIOBASIC USER MANUAL Models 210 310 410...

Страница 2: ...recommend that you read this user manual through thoroughly before using the cabinet Gram Commercial A S does not guarantee safety if the appliance is used for anything other than its intended use Co...

Страница 3: ...tion 10 Voltage free contact 11 Connecting to electricity 12 Start up 14 The digital display controls 15 General introduction to the controller interface 16 Alarm settings 17 Buttons and useful shortc...

Страница 4: ...s in Category 3 Zone 2 locations where gas atmospheres may be present Temperature setpoint range and ambient requirements model temperature setpoint range Minimum ambient operating temperature Maximum...

Страница 5: ...BIOBASIC MANUAL 5 Symbols used Hazard Risk of electric shock Risk of material damage Risk of personal injury Risk of burning freezing Info...

Страница 6: ...irect sunlight or close to heat sources Avoid placement of the cabinet in a chloric acidic environment due to risk of corrosion Clean the cabinet with a mild soap solution prior to use If the cabinet...

Страница 7: ...hould be levelled as shown in the illustrations to the left For cabinets equipped with castors the floor must be level to ensure stable positioning and safe use When the cabinet is positioned the 2 fr...

Страница 8: ...tip over when the drawers are drawn to the outermost position or the door is open Brackets for securing the cabinet are included from the factory Find the instructions for the tilt bracket below The...

Страница 9: ...mounting Wall mounting brackets can be supplied if specified Allowing the cabinet to be lifted from the floor Find instructions on wall mounting of a BIOBASIC 210 below the same procedure applies for...

Страница 10: ...visual inspection of the cabinet must be conducted prior to putting the cabinet into service Check the cabinets structural integrity that door frames and doors don t have deformities gaskets seal pro...

Страница 11: ...ctively Common NO Normally Open and NC Normally Closed The moment voltage is applied the controller draws the relay this makes it possible for the controller to respond to both high and low temperatur...

Страница 12: ...e zone If the appliance is to be installed in a zone 2 environment specialist personnel should perform the installation or be consulted beforehand in order to ensure that the appliance is installed in...

Страница 13: ...iance must be moved to an area in which there is no risk of ignition caused by the electrical components or gases contained in the appliance Never use the cabinet if the plug is damaged The cabinet sh...

Страница 14: ...13 6 8 14 1 On Off 2 Select or confirm a menu parameter 3 Navigate upwards in a given menu raise a given value 4 Navigate downwards in a given menu lower a given value 5 Parameter settings menu go a m...

Страница 15: ...P DOWN SELECT ON OFF MENU BACK UP DOWN SELECT 10 9 5 3 2 1 15 4 9 Key pad lock engaged 10 Display 11 Temperature and or door alarm registered 12 Parameter settings menu is open 13 Defrost in progress...

Страница 16: ...tance after a power outage or when plugging the cabinet in for the first time Setpoint temperature adjustments Temperature adjustments are done by pressing for 3 seconds prompting the set point temper...

Страница 17: ...d A3 1 on 0 off Alarm settings Buttons and useful shortcuts Buttons Press for Function Acknowledge alarm mute for 5 minutes 3 seconds Access the alarm settings 5 seconds Access the parameter settings...

Страница 18: ...on or off tEr Relay test Component test tC C Test compressor relay tF Min Test evaporator fan td Min Test defrost element relay RF models tA Min Test alarm relay will trip voltage free tdP On off Test...

Страница 19: ...refrigeration system will use an emergency program to make the cabinet run Temperature stability will be affected Service is required F2 Error on the evaporator sensor Service is required Please find...

Страница 20: ...sensor is offset if the actual temperature in the cabinet does not match the setpoint despite taking the hysteresis into consideration Offset of A sensor is named CA Offsetting the A sensor Press and...

Страница 21: ...so that the refrigeration system stops 1 0K before and starts 1 0K later than the setpoint normally otherwise would dictate Example 2 The temperature in the cabinet is operating warmer than the actua...

Страница 22: ...its components Items placed on the bottom of the cabinet will cause the air circulation to be impeded reducing the cabinets performance Items should be evenly distributed in the cabinet with minimum l...

Страница 23: ...careful not to damage the defrost water tube and heating element located in the tray when cleaning Cleaning agents containing chlorine or compounds of chlorine as well as other corrosive agents may n...

Страница 24: ...dity in the storage chamber iced evaporator and thus reduced cooling capacity and in some cases decreased lifetime expectancy of the cabinet It is therefore very important to be aware of the door gask...

Страница 25: ...eplaced with original parts from Gram Commercial A S Gram Commercial A S can only guarantee functional and safety requirements on the cabinets if above mentioned is adhered to The cabinets refrigerati...

Страница 26: ...upwards 3 Dismantle the hinge at pos A and lift off the door 4 Dismantle the hinge at pos B and mount it at pos F 5 Turn the door 180 and fix it at the hinge pos F 6 Mount the hinge from pos A in pos...

Страница 27: ...can lead to unexpected movements of the cabinet Lock the castors after installation Negligence of these precautions can lead to injuries The re evaporation tray re evaporation tray heating element pre...

Страница 28: ...of properly Products that are labelled with a crossed out wheelie bin is electric and electronic equipment The crossed out wheelie bin symbolizes that waste of this type can not be disposed of with un...

Страница 29: ...Electrical equipment for measurement control and laboratory use EMC requirements General requirements DS EN 60079 0 2012 Electrical apparatus for explosive atmospheres DS EN 60079 11 2012 Explosive a...

Страница 30: ...30 BIOBASIC MANUAL Wiring diagrams...

Страница 31: ...BIOBASIC MANUAL 31...

Страница 32: ...32 BIOBASIC MANUAL Piping diagram...

Страница 33: ...BIOBASIC MANUAL 33 Notes...

Страница 34: ...34 BIOBASIC MANUAL Notes...

Страница 35: ...BIOBASIC MANUAL 35...

Страница 36: ...Gram Commercial A S Aage Grams Vej 1 DK 6500 Vojens Tel 45 7320 1300 Info gram bioline com Item number 765041746 Rev 180518 Language English...