1137-INS-LAB-RevE19
16
MAINTENANCE
WARNING: Ensure device is unplugged before maintenance.
WARNING: Do not disassemble or modify the device or cuff.
s
NOTICE: Do not drop the device. It is not shock proof.
s
NOTICE: Do not twist the cuff or bend the tube.
s
NOTICE: Do not store the device or cuff in direct sunlight, high
temperature, moisture, or severe dust.
s
NOTICE: Do not use thinner, benzene, or other harsh cleaners
to clean the device.
s
NOTICE: Do not wash or immerse any part of the device.
s
NOTICE: Remove all batteries if the device is not likely to be
used for a long period of time.
CLEANING
1. Moisten a clean, soft cloth with a solution of 5-10% mild detergent to
clean water. Gently wipe the device with a contact time of at least 10
minutes.
2. Use a clean, soft cloth moistened with clear, cool water to remove
any chemical residue remaining on the device.
3. If necessary, repeat steps 1 and 2 to ensure the device is clean.
4. Wipe dry with a clean, soft, cloth to avoid water stains.
DISINFECTION (WHEN CLEANING PROCESS IS COMPLETE)
1. Use 70 to 75% isopropyl alcohol in a spray bottle and spray both
sides of the cuff with it, with a contact time of at least 10 minutes.
2. Use a clean, soft cloth moistened with clear, cool water to remove
any chemical residue remaining on the device.
3. Air dry the device and cuff at a room temperature of 68°F ~ 86°F
(20°C ~ 30°C) for 3 ~ 4 hours.
STORAGE
s
NOTICE: When the device is often in use, store it in a dry,
clean, and dust-free environment at a moderate temperature of
41°F ~ 104°F (5°C ~ 40°C), < 93 % RH. If the device is not likely
to be used for a long period of time, store it at a temperature of
-4°F ~ 131°F (-20°C ~ 55°C), < 93 % RH.
s
NOTICE: Remove all batteries if the device is not likely to be
used for a long period of time.