FR601-INS-LAB-RevC20 18

Heat and Massage Unit Troubleshooting Guide

The cables and connections need to be validated. This will require the services of a 
maintenance person and a clear working area.
1.  INSPECTION: Perform the following checks. If a working recliner of the same model is 

available, use it for comparison.
a)  Is the AC cord properly connected to the power supply? Is the cable from the power 

supply to the control module junction box in place?

b)  With the power supply AC cord plugged into a live AC outlet, does the small light to 

the left of the white seat symbol (located on the switch) light up when the switch is 
pressed repeatedly or is there any indication of massage operation?

  WARNING: Disconnect the power cord from the AC outlet before performing the 

following steps!

c)  Is there any sign of external damage to the handheld pendant or wire?
d)  Is there any sign of external damage to the recliner?
e)  Do any of the cables show any sign of damage or pinch points?
f)  Are all cable connections to the control module in place? Are the connectors fully 

engaged?

g)  Are any of the connectors damaged?
h)  Are the wires within the connectors attached to the pins? Each wire may have to be 

tugged lightly to check if the wire is still crimped to the pin.

4. POST-INSPECTION:

a)  If improper or loose connections are found, reconnect them firmly (unless a loose 

wire has been found). Plug the power supply AC cord into a live wall socket and 
operate the unit through all functions.

b)  If the handheld pendant or wire is damaged, notify Graham-Field so that a handheld 

pendant can be sent for replacement.

c)  If damage to any other electrical cable or component, or disconnected wires are 

found, please notify Graham-Field.

5. ACTIONS:

a)  If full operation is restored, please advise Graham-Field of your successful outcome.
b)  If the fault cannot be corrected by simple re-connection, please inform Graham-Field 

of the results of your inspection in detail so that we may determine the appropriate 
action to be taken.

If you have any questions about the setup and / or operation of the Heat and Massage Unit, 
call the Graham-Field Technical Support Group at 1.770.368.4700.

Содержание Lumex FR601 Series

Страница 1: ...at and Massage Function Assembly and Operating Instructions IMPORTANT READ THIS MANUAL BEFORE OPERATING YOUR LUMEX FR601 FR601H SERIES CLINICAL CARE RECLINER WITH PIVOT ARM SAVE THIS MANUAL FOR FUTURE...

Страница 2: ...SITION 10 PIVOT ARM OPERATION 11 OPEN THE PIVOT ARM 11 CLOSE THE PIVOT ARM 11 FOOTPLATE OPERATION 12 HEAT AND MASSAGE UNIT SETUP INSTRUCTIONS GEN 3 PENDANT OPERATED 13 INSTALLATION AND OPERATING GUIDE...

Страница 3: ...ilar to the following or specific to the operations Please note the following special statements used throughout this manual and their significance WARNING Indicates a potential hazard situation or un...

Страница 4: ...us condition which may result in serious injury and will void the warranty WARNING This product can expose you to chemicals including Di 2 ethylhexyl phthalate DEHP which is known to the State of Cali...

Страница 5: ...T DESCRIPTION upholstered back pivot arms upholstered seat upholstered legrest skirt upholstered footrest self storing side table frame frame Trendelenburg activation pedal pivot arm release push hand...

Страница 6: ...hanism 2 Remove two pre installed threaded knobs from back one on each side 3 Carefully guide the upholstered back onto the mechanism inserting both bayonet links at the same time through the two cuto...

Страница 7: ...ures three total locking casters identified by black actuator tabs that lock both the wheel and the swivel To activate the total lock simply press down on the BLACK actuator tab until it locks To rele...

Страница 8: ...disengaged position Side Table See illustration above for side table operation Raise Side Table Raise the side table all the way up and slide in toward the armrest until tabs engage slots in mount br...

Страница 9: ...er hand Push back on push handle to recline To return grasp push handle and pull forward to raise the recliner s back s NOTICE DO NOT use the footrest by itself to return the recliner to an upright po...

Страница 10: ...he other With your foot push down on the Trendelenburg activation pedal either side of the recliner while gently pushing down with your hand on the recliner back until it comes to a stop To bring recl...

Страница 11: ...1 Pivot the Arm back toward the front of the recliner until the pivot arm stops Step down on the YELLOW RELEASE LEVER located toward the front of the recliner Continue to push the pivot arm over the...

Страница 12: ...can now rest their feet on the footplate while being transported To return the footplate to the stored position push the footplate back underneath the recliner until it stops and is safely stored und...

Страница 13: ...t change the system wiring in any way WARNING If components are damaged or missing contact your Graham Field authorized distributor immediately DO NOT use substitute parts Use only Lumex replacement p...

Страница 14: ...not in use or for caregiver use place the pendant in the pendant storage bracket mounted on the back of the recliner as shown above All Heat and Massage functions are controlled by the handheld penda...

Страница 15: ...is will permit uninterrupted access to the back compartment while making the final cable connection 3 Remove the handheld pendant and power cord from the back compartment Place the handheld pendant in...

Страница 16: ...operated using the handheld pendant shown above Faceplate Selector Switch Heat and Massage Handheld Pendant 3 The Pendant has two membrane switches The left switch controls the Massage Function and th...

Страница 17: ...elector switch for Setting 1 High Heat 104 F 40 C brightest LED b Press selector switch AGAIN for Setting 2 Medium Heat 101 F 38 6 C bright LED c Press selector switch AGAIN for Setting 3 Low Heat 100...

Страница 18: ...f Are all cable connections to the control module in place Are the connectors fully engaged g Are any of the connectors damaged h Are the wires within the connectors attached to the pins Each wire may...

Страница 19: ...staining Clean area with kerosene or naphtha WARNING When using kerosene or naphtha do so in a well ventilated area and keep it away from fire flame and recliner occupant Waxing and refinishing impro...

Страница 20: ...of S C Johnson Son Inc Spradling International is not sponsored by affiliated with or endorsed by such trademark owners CDC RECOMMENDATIONS CLEANING AND DISINFECTING STRATEGIES FOR ENVIRONMENTAL SURFA...

Страница 21: ...sinfectant fogging in patient care areas Avoid large surface cleaning methods that produce mists or aerosols or disperse dust in patient care areas Follow proper procedures for effective uses of mops...

Страница 22: ...3951 DB MAARN The Netherlands 31 0 6 516 536 26 www cepartner4u com TECHNICAL SPECIFICATIONS Overall Height 51 1 2 132 cm Overall Width 34 1 2 88 cm Overall Depth 38 97 cm Width Between Arms 25 63 5...

Страница 23: ...ut not limited to filters fuses gaskets lubricants and charts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement parts installation or other work perform...

Страница 24: ...rmation can be found on our website www grahamfield com 2019 GF Health Products Inc All Rights Reserved Graham Field and Lumex are trademarks of GF Health Products Inc GF Health Products Inc is an ISO...

Отзывы: