Heat and Massage / Heat Only Unit Troubleshooting Guide

The cables and connections need to be validated. This will require the services of a maintenance 
person and a clear working area.
1.  INSPECTION: Perform the following checks. If a working recliner of the same model is available, 

use it for comparison.
a)  Is the AC cord properly connected to the power supply? Is the cable from the power supply 

to the control module junction box in place?

b)  With the power supply AC cord plugged into a live AC outlet, does the small light to the 

left of the white seat symbol (located on the switch) light up when the switch is pressed 
repeatedly or is there any indication of massage operation?

  WARNING: Disconnect the power cord from the AC outlet before performing the following 

steps!

c)  Is there any sign of external damage to the handheld pendant or wire?
d)  Is there any sign of external damage to the recliner?
e)  Do any of the cables show any sign of damage or pinch points?
f)  Are all cable connections to the control module in place? Are the connectors fully engaged?
g)  Are any of the connectors damaged?
h)  Are the wires within the connectors attached to the pins? Each wire may have to be tugged 

lightly to check if the wire is still crimped to the pin.

2. POST-INSPECTION:

a)  If improper or loose connections are found, reconnect them firmly (unless a loose wire has 

been found). Plug the power supply AC cord into a live wall socket and operate the unit 
through all functions.

b)  If the handheld pendant or wire is damaged, notify Graham-Field so that a handheld 

pendant can be sent for replacement.

c)  If damage to any other electrical cable or component, or disconnected wires are found, 

please notify Graham-Field.

3. ACTIONS:

a)  If full operation is restored, please advise Graham-Field of your successful outcome.
b)  If the fault cannot be corrected by simple re-connection, please inform Graham-Field of the 

results of your inspection in detail so that we may determine the appropriate action to be 
taken.

If you have any questions about the setup and / or operation of the Heat and Massage / Heat Only 
Unit, call the Graham-Field Technical Support Group at 1.770.368.4700.

Содержание LUMEN Ortho-Biotic II FR597G Series

Страница 1: ...s with Heat and Massage Series FR597PH Ortho Biotic II Recliner with Fully Upholstered Arms with Heat and Massage Series FR597GHO Ortho Biotic II Recliner with Wooden Arm Caps with Heat Only Series FR...

Страница 2: ...POSITION 8 FOOTPLATE OPERATION 9 UNIVERSAL HEADREST OPERATION 9 OPTIONAL ACCESSORIES 10 OPTIONAL SIDE TABLE 10 OPTIONAL SIDE TABLE OPERATION 10 OPTIONAL IV POLE MOUNT 10 RECLINERS WITH HEAT AND MASSAG...

Страница 3: ...re additional safety statements in other chapters or sections which may be the same as or similar to the following or specific to the operations Note the following special statements used throughout t...

Страница 4: ...ARNING This product can expose you to chemicals including Di 2 ethylhexyl phthalate DEHP which is known to the State of California to cause cancer and birth defects or other reproductive harm For more...

Страница 5: ...ay to protect the casters during shipment For shipment the recliner is covered with a carton as well as a topper for protection The topper carton and tray are strapped to a pallet for shipment as show...

Страница 6: ...II RECLINER DESCRIPTION AND FEATURES Directional Locking Caster 1 Total Locking Caster 3 Trendelenburg Foot Pedal RIGHT SIDE ONLY Pull Out Footplate Legrest Recline Adjustment Lever Padded Push Handl...

Страница 7: ...caster The fourth caster located on the right rear position of the recliner is a directional swivel locking caster only Identified by a green actuator tab the directional lock caster will lock in one...

Страница 8: ...patient up and return to the upright position TRENDELENBURG SHOCK POSITION Info Trendelenburg shock position can ONLY be activated by a caregiver Trendelenburg Position CANNOT be activated by a patien...

Страница 9: ...c straps until the straps are taut Secure the pillow in position by pressing the hook and loop sections of the elastic straps together Once the patient is positioned in the recliner adjust the positio...

Страница 10: ...in mounting bracket Press down lightly on table to ensure locking tabs are engaged Lowering Side Table Lift the side table up slightly until tabs disengage slots and pull table away from armrest then...

Страница 11: ...of this product Description The Lumex Healthcare Seating Heat and Massage Heat Only Unit is a low voltage 12V Heat and Massage Heat Only Unit controllable by the user and or caregiver via a handheld p...

Страница 12: ...ly Installation Instructions 1 Place the back cushion over the vertical bayonet lugs of the recliner mechanism Secure the back on each side with the threaded screw knobs per the previous Upholstered B...

Страница 13: ...tabs snap into place as shown above Connector Security Lock 5 If you need to replace the Hand Pendant remove the small white connector security lock as shown above 6 Remove the tape holding the back...

Страница 14: ...200RevA20 SEE OPERATING MANUAL FOR COMPLETE INSTRUCTIONS HIGH HEAT PRESS 1 TIME MEDIUM HEAT PRESS 2 TIMES LOW HEAT PRESS 3 TIMES 3 The Heat and Massage Pendant Selector Switch Faceplate shown at above...

Страница 15: ...The Heat Function Settings are as follows a Press selector switch for Setting 1 High Heat 104 F 40 C brightest LED b Press selector switch AGAIN for Setting 2 Medium Heat 101 F 38 6 C bright LED c Pre...

Страница 16: ...s f Are all cable connections to the control module in place Are the connectors fully engaged g Are any of the connectors damaged h Are the wires within the connectors attached to the pins Each wire m...

Страница 17: ...h a clean damp cloth to get rid of any excess soap Repeat as needed and then let dry For Guardian Patriot Plus Renaissance and Savoy fabrics only If there is any build up that can t be removed with so...

Страница 18: ...s and devices or any environmental surfaces such use is counter to label instructions for these toxic chemicals Follow manufacturers instructions for cleaning and maintaining noncritical medical equip...

Страница 19: ...op head at the beginning of each day and also as required by facility policy or after cleaning up large spills of blood or other body substances Clean mops and cloths after use and allow to dry before...

Страница 20: ...ic types code and fiber content to aid sorting of plastic parts Product Metal Scrap Cable Scrap Electronic Scrap Plastic Recycling or Combustion FR597G X X N A X FR597GH X X X X FR597GHO X X X X FR597...

Страница 21: ...ndelenburg Position 74 188 cm Back Angle Fully Upright 120 With Legrest Extended 139 Fully Reclined 180 Back Height above Seat 31 78 4 cm Seat Height 21 53 3 cm Seat Width 23 58 4 cm Seat Depth 20 5 5...

Страница 22: ...nature including but not limited to filters fuses gaskets lubricants and charts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement parts installation or...

Страница 23: ...urning to stored position 8 Limited warranty 22 N NOTICE statement significance 3 O Operation 7 P Packaging 5 R Recline operation 8 Recliner operation 7 Reclining by caregiver 8 Reclining by patient 8...

Страница 24: ...2014 GF Health Products Inc All Rights Reserved Lumex and Ortho Biotic are trademarks of GF Health Products Inc Tente is a trademark of TENTE GmbH Co FED REP GERMANY GF Health Products Inc is an ISO...

Отзывы: