7  TROUBLESHOOTING

 

WARNING:

 Service and repair of the Gendron 3648 Bariatric Bed should be performed 

only by a GF authorized distributor. GF is not responsible for any consequence resulting 

from any unauthorized service or repair.

The following list of encountered problems and their solutions will assist you in determining what 
may be causing your 3648 Bariatric Bed not to function as designed. If a problem occurs which is 
not listed below, please contact your GF authorized distributor or GF Technical Support for help. 
Do not attempt to repair 3648 Bariatric Bed components or parts, as this may invalidate your 
warranty or cause further problems.

Problem

Symptom

Solution

No Power

1. Is green light on control 

module illuminated?

If yes go to Power but no function section.

2. Is the mains cable damaged? If yes, replace mains cable.

3. Is it plugged into the wall?

Plug mains cable into a functional wall receptacle.

4. Is it plugged into control 

module?

Plug mains cable into the control module.

5. Is there power to the wall 

receptacle?

If yes, replace control module.

Power but 

no function

1. Is green light on control 

module illuminated?

2. Is there damage to the 

handset or cable?

If yes, replace handset.

3. Is the handset plugged in?

Plug in handset.

4. Are all of the motor cables 

plugged into the actuators and 

control module?

Plug motor cables into actuators and control 

module.

5. Are any of the actuator motor 

cables damaged?

If yes, replace the damaged motor cable.

6. Is the control module 

chirping?

If yes, call GF Tech Support.

Actuator 

not 

functioning

1. Is motor cable plugged into 

actuator and control module?

Plug motor cable into actuator and control module.

2. Is the motor cable damaged? If yes, replace motor cable.

3. Is the actuator damaged?

If yes, replace actuator.

4. Does the actuator make any 

sounds when function button is 

pushed?

If yes, check for any mechanical binding on bed.

5a. Can you change ports on 

the control module and get it to 

work? (Ports 1 and 3 or 2 and 4)

If initial suspect actuator still does not move but 

switched port actuator does, then replace the motor 

cable and suspect actuator. If suspect actuator 

moves but switched port actuator does not, then 

replace the control module.

5b. Does other actuator that the 

port was switched with not work 

now?

 

WARNING:

 If you experience a problem with the 3648 Bariatric Bed and are unable to 

service it yourself, contact your GF authorized distributor.

GD_3648-INS-LAB-RevA20 20

Содержание GENDRON 3648 Series

Страница 1: ...B RevA20 Read this manual before operating your 3648 Bariatric Bed Save this manual for future use The most current version of this manual can be found online at gendroninc com 3648 SERIES BARIATRIC B...

Страница 2: ...ES 8 4 ASSEMBLY AND SETUP INSTRUCTIONS 9 FEATURES AND COMPONENTS 9 ASSEMBLY INSTRUCTIONS 10 SETUP INSTRUCTIONS 11 DISASSEMBLY FOR STORAGE OR TRANSPORT 12 5 OPERATION INSTRUCTIONS 13 FUNCTION AND BASIC...

Страница 3: ...health professional can determine the suitability of this bed Contraindications WARNING Use this device only as intended and described in this manual Do not use attachments or accessories not recomme...

Страница 4: ...er to CARE AND MAINTENANCE section of this manual Maintenance MUST be performed by qualified personnel ONLY SAVE THESE INSTRUCTIONS SIGNIFICANCE OF SAFETY STATEMENTS Please note the following special...

Страница 5: ...tributed over the surface of the bed Do not lie sit or lean in such a way that your entire body weight is placed only on the raised head or foot section of the bed This includes when assisting the use...

Страница 6: ...NING Check all parts for shipping damage before using In case of damage DO NOT USE the equipment Contact the carrier or your GF authorized distributor for further instructions WARNING This product can...

Страница 7: ...equipment can affect medical electrical equipment This Product is intended for use in the electromagnetic environments specified 2 The end user of this product should assure it is used in such an env...

Страница 8: ...tainers and packing material for use in storing until use of Bed is required Inspection 1 Examine the exterior of the 3648 Bariatric Bed for nicks dents scratches or other damage inspect all component...

Страница 9: ...ASSEMBLY AND SETUP INSTRUCTIONS FEATURES AND COMPONENTS Brake Caster Head Board Back Rest Side Rail Foot Board Steer Caster Leg Lift Side Rail Release Pin IV Socket Lift Handle GD_3648 INS LAB RevA20...

Страница 10: ...above left 2 Install the footboard posts in the square sockets at the foot end with the board away from the mattress as shown at above right 3 Install the side rails as shown at above left 4 Install t...

Страница 11: ...the round bar on the head end frame 3 Swing the lift handle out and slowly lower the bed Keep both hands on handgrips until the bed has been completely lowered to reduce the risk from pinch areas 4 I...

Страница 12: ...th actuator cables from foot end of frame to head end of frame 4 Remove quick release pins on both sides of bed frame 5 Swing lift handle out and slowly lift bed Keep both hands on handgrips until bed...

Страница 13: ...ing members of the bed Ensure the control module and the handset cords are properly positioned before operating the bed Do not allow either of these cords to get caught in the moving members of the be...

Страница 14: ...led retain pivot pins and hairpin clip Control Module The Control Module is designed to control all actuators in the system The Control Module mounts directly on one of the actuators as shown Handset...

Страница 15: ...e and lower head section 2 Raise and lower foot and leg section and 3 Raise and lower bed height UPRIGHT BACK POSITION WARNING Before operating the backrest head section always ensure that casters are...

Страница 16: ...veling and allow better forward and backward movement The bed can now be moved with greater stability Push the bed from the head end Do not lock the other steering caster Using the Four Caster Free Fu...

Страница 17: ...who get out of bed and walk unsafely without assistance For at risk patients use only with proper bed side rail protective barriers There is also an increased risk of entrapment when bed rails are use...

Страница 18: ...ous condition which may result in serious injury and will void the warranty WARNING The bed s electronic system contains no serviceable parts Do not open the Motors Cables Motor Cord or Handset Only a...

Страница 19: ...proved lubricant X Ensure release pins spring back to locked position X Ensure retaining pins are properly installed in clevis pins X Ensure all electrical cables are free of frays kinks and deteriora...

Страница 20: ...odule illuminated 2 Is there damage to the handset or cable If yes replace handset 3 Is the handset plugged in Plug in handset 4 Are all of the motor cables plugged into the actuators and control modu...

Страница 21: ...28 6 cm with headboard and footboard installed Deck Width 48 121 9 cm fixed width Deck Length 84 213 4 cm Deck Height 15 to 32 38 1 cm to 81 3 cm above floor Weight 371 lb 168 3 kg Casters four total...

Страница 22: ...n durable nature including but not limited to casters filters fuses gaskets lubricants and charts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement part...

Страница 23: ...nd wire 14 H Handling procedures 8 Handset 14 Handset 3 function 15 Handset operation 15 I Important safety instructions 4 Info statement significance 4 Inspection 8 Inspection and maintenance 19 Inst...

Страница 24: ...tion can be found on our website grahamfield com gendroninc com 2020 GF Health Products Inc All Rights Reserved Gendron is a trademark of GF Health Products Inc GF Health Products Inc is an ISO 13485...

Отзывы: