5.
Dispense about 300–500 cc of material into a
graduated cylinder. The amount the system
measures will display in the measured volume
field
.
6.
Press
to end the calibration.
7.
Press
to get to the dispensed volume field
, then press
to enter the field. Enter
the amount of material in the cylinder.
8.
After the volume is entered, the system calculates
the new k-factor and shows it on Setup Screen 4.
NOTE:
NOTE:
NOTE: To clear the counter and begin the
calibration again, press
, move briefly to
another screen, then return to Setup Screen 4
and start over. If you press
without leaving
the screen, the counter will continue from where
it is, without clearing.
9.
Press
to exit the screen.
10. Press
to exit Setup Mode.
Setting
Setting
Setting Modbus
Modbus
Modbus Address
Address
Address
See Setup
Setup
Setup Screen
Screen
Screen 5
5
5. By default, the Modbus is
set to Off
. If you need the Modbus, set the
Modbus mode to SLAVE
. The address
value is between 1 and 247. The modbus address
corresponds to the address of the Informer. See
Appendix A for further information.
Update
Update
Update Software
Software
Software
Software updates are installed using a software
token (PN 16P468), which is sent automatically when
a new version of the software is released. Manual
3A1244 will accompany any necessary software
updates. Follow all instructions and warnings in
Manual 3A1244 to update your Informer software.