86
GMC-I Messtechnik GmbH
8.3
Query Commands for Status and Events Management
1)
When using RS232 interface
without
IEEE488 interface, value 1 is generally returned as a response.
2)
When using RS232 interface
without
IEEE488 interface, the invalid value 127 is generally returned as a response.
Read out via interface Response string for remote operation (example):
Resp.
string
length
Explanation
in Chapter
on Page
RS 232C
IEEE 488
Qu
er
y S
tatus and E
vent Re
gisters
CRA?
Condition register A query
X
X 002
3
D7: SEQB
Status message: sequence function active (run, hold)
D6: TRGA
Signal at analog interface trigger input
D5: OTPA
Overtemperature (excessive temperature)
D4: OVPA
Overvoltage protection triggered
D3: S12A
SIG1_OUT/SIG2OUT active
D2: OL
Overload
D1: CCR
Output in current regulating mode
D0: CVR
Output in voltage regulating mode
ERA?
Device dependent event register A query
X
X 032
3
D7: SEQI
Sequence function completed (inactive) (ready)
D6: OTPI
Ready for operation message after OTPA
D5: OTPA
Overtemperature message
D4: OVPA
Overvoltage protection triggered
D3: OCPA
OCP function triggered:
D2: OL
Overload has occurred
D1: CCR
Current regulation has occurred
D0: CVR
Voltage regulation has occurred
ERAE?
Device dependent event register A enable query
X
X 032
3
ERB?
Device dependent event register B query
X
X 128
3
D7: TCE
Self-test error or error during adjustment
D6: TRGA
Signal at analog interface trigger input
D5: SEQE
Sequence function error message
D4: OUTE
Error message after OUTPUT ON, disabling
via trigger input at analog interface
D3: DDTE
Define device trigger function error message
D2: LIME
Limit error, ULIM < USET or ILIM < ISET
D1: S2A
SIG2_OUT, active signal occurred
D0: S1A
SIG1_OUT, active signal occurred
ERBE?
Device dependent event register B enable query
X
X 128
3
*
ESE?
Standard event status enable query
X
X 032
3
*
ESR?
Standard event status register query
X
X 144
3
D7: PON
Power On
D6: 0
(URQ user request)
D5: CME
Command error
D4: EXE
Execution error
D3: 0
(DDE device depend error)
D2: QYE
Query Error
D1: 0
(RQC request control)
D0: OPC
Operation complete
*
IST?
Individual status query
X
1)
X 0
1
*
OPC?
Operation complete query
X
X 1
1
*
PRE?
Parallel poll enable register enable query
X
X 012
3
*
PSC?
Power-on status clear query
X
X 0
1
*
SRE?
Status request enable query
X
X 012
3
*
STB?
Read status byte query
X
2)
X 034
3
D7: 0
D6: RQS MSS
D5: ESR
Standard event register
D4: MAV
Message Available
D3: ERA
Event register A
D2: ERB
Event register B
D1: 0
D0: 0
Содержание SSP 120 Series