GMC-I Messtechnik GmbH
75
ULIM, ULIM? – Voltage Setting Limit Value
Function and manual operation:
See explanation on page 25.
Programming
Setting command:
UL
IM
v
Parameter:
v (value)
Parameter type:
Real number
Setting range:
0
v
ULIM
max
l
Default setting:
after RESET (*RST)
nominal voltage
Example
(HP BASIC):
OUTPUT 713;"ULIM 20"
{voltage setting range limit: 20 V}
Comment
ULIM may not be set to a value of less than USET!
Any setting command to this effect is not executed.
Bit 1
would otherwise be set in event register B (
limit error
), and
bit 4
in the standard event register (
execution error
).
If the selected value is not a whole number multiple of the step
size, it is rounded off accordingly.
If necessary, the numeric value is rounded off once again for
the 4-place digital display.
The
ULIM
function does not effect the
USET
setting by means of
a control signal via the analog interface.
Query
Displays the selected final value for voltage setpoint USET.
Programming
Query command:
UL
IM
?
Response string:
ULIM
v
Length:
13 characters, fixed
Parameter
v (value)
Format
v
: +nnn.nnn
Example
(HP Basic):
OUTPUT 713;"
UL
IM
?
"
ENTER 713; A$
DISP A$
Display: ULIM
+020.000
UMAX? – Maximum Measured Voltage Value
Function and manual operation:
See explanation in chapter 4.8.1.
Programming
Query command:
UM
AX
?
Response string:
UMAX
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nnn.nnn
Example
(HP Basic):
OUTPUT 713; "UMAX?"
ENTER 713; A$
DISP A$
Display:
UMAX +028.125
UMIN? – Minimum Measured Voltage Value
Function and manual operation:
See explanation in chapter 4.8.1.
Programming
Query command:
UM
AX
?
Response string:
UMAX
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nnn.nnn
Example
(HP Basic):
OUTPUT 713; "UMIN?"
ENTER 713; A$
DISP A$
Display:
UMIN +027.350
UOUT? – Querying the Momentary Voltage Value
Function and manual operation:
See explanation on page 23.
Programming
Query command:
UO
UT
?
Response string:
UOUT
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nnn.nnn
Example
(HP Basic):
OUTPUT 713; "UOUT?"
ENTER 713; A$
DISP A$
Display:
UOUT +027.350
Device Type
Nom. voltage [V]
20
40
80
360
ULIM
max
[V]
20
40
80
360
Step size [mV]
1.0
1.0
1.0
1.0
Содержание SSP 120 Series
Страница 2: ...2 GMC I Messtechnik GmbH ...