82
GMC-I Messtechnik GmbH
Classification Parameter – VDE 0701-0702-VLTG
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the
basis of length and cross-section (length only in the case of EL1). Data
remain in memory until a new entry is made.
Classification Parameter – VDE 0701-0702-PRCD
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
3
New classification parameters “PRCD type”
(only displayed if “Standard
VDE 0701-0702-PRCD” parameter is selected):
–
PRCD (standard):
For the testing of simple circuit breaker safety adapters in
which the protective conductor is permanently connected.
–
PRCD (SPE):
(SPE = switched protective earth) for testing PRCDs in which
the protective conductor is only connected in switched-on
condition.
–
PRCD-S (SPE):
For testing type PRCD-S circuit breaker safety adapters.
–
PRCD-K (SPE):
For testing type PRCD-K circuit breaker safety adapters.
Note
The standard or standard variant associated with the
respective selector switch position corresponds to the
default setting.
Ax means that the standard variant, VDE 0701-0702-
PRCD, can be selected in each of the preset switch posi-
tions.
Note
For more information on testing single-phase and 3-
phase type S and K PRCDs by simulating faults see
“
PROFITEST PRCD
test adapter” on our website.
Note
Testing of PRCDs (test sequences and time to trip) is only
possible for DUTs with a nominal voltage of 230 V.
Parameter
Setting Options / Meaning
1/2
Standard
VDE 0701-0702, see table below
VDE 0701-0702-EDV, see table above
IEC 60601, 2
nd
edition, see table above
EC 60601, 3
rd
edition, see table above
EC 60601, 3
rd
edition GPA, see table above
IEC 62353, see previous table
VDE 0701-0702-VLTG
VDE 0701-0702-PRCD, see following table
IEC 60974-4, see table below
Protection class
1, 2
PCI
Connection type
1, 2
Test socket
Adapter: AT3-IIIE
Adapter: EL1 adapter
Adapter: VL2E adapter
2/2
Measurement type (MT)
1
VLTG
2
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & measurement type
Protection class only (PC)
Measurement type only (MT)
Disabled: No auto-detection:
All classification parameters such as connection, protection
class and measurement type must be entered manually.
Parameter
Setting Options / Meaning
1/2
Standard
VDE 0701-0702, see table below
VDE 0701-0702-EDV, see table above
IEC 60601, 2
nd
edition, see table above
EC 60601, 3
rd
edition, see table above
EC 60601, 3
rd
edition GPA, see table above
IEC 62353, see table above
VDE 0701-0702-VLTG, see previous table
VDE 0701-0702-PRCD
2
IEC 60974-4, see following table
Protection class
1, 2
PCI, PCI+II
Connection type
1, 2
Test socket
2/2
Measurement type (MT)
1
PRCD
3
PRCD type
3
PRCD (standard)
PRCD (SPE)
PRCD-S (SPE)
PRCD-K (SPE)
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & measurement type
Protection class only (PC)
Measurement type only (MT)
Disabled: No auto-detection:
All classification parameters such as connection, protection
class and measurement type must be entered manually.