68
Gossen Metrawatt GmbH
15.9
Probe Voltage – U
Mains to Test Socket
Schematic Diagram
Wiring Diagram
Permanently Connected DUT
Schematic Diagram
Wiring Diagram
Direct, alternating and pulsating voltages of up to 253 V can be
measured. Two connection types are available, one of which has
to be selected in the parameters menu.
Setting Measuring Parameters for U
Probe
Test Procedure
➭
Set the rotary switch to the
U
position.
➭
Connect the DUT’s mains plug to the test instrument’s test
socket.
➭
Start the test:
Press the
START
/STOP
key.
➭
PE-P1 (with mains):
Acknowledge the warning which in-
dicates that line voltage will be connected to the test
socket.
➭
Switch the device under test on.
➭
Contact the ungrounded output for safety extra-low voltage
with test probe P1.
➭
Polarity can be set via direct selection immediately be-
fore measurement is started, without having to switch to
the parameters menu.
➭
The measured values are displayed. The measured
value recording icon shown at the right appears. Each
time this key is pressed, the currently displayed mea-
sured value is saved to buffer memory.
➭
Turn off the device under test.
➭
End the test:
Press the
START
/STOP
key.
The save icon appears (floppy disk showing the num-
ber of measured values stored to buffer memory) and
prompts you to save the measured values to a test ob-
ject ID number.
➭
Press the
ESC
key in order to discard the measured
values stored to buffer memory and acknowledge by
pressing the key shown at the right.
Single Measurements,
Rotary Switch Level: Green
Ro
ta
ry
Switc
h
Po
si
ti
o
n
Meas. Type
With Mains to
Test Socket
Meas. Type
Without
Mains to Test
Socket
Measuring Functions
U
PE - P1
U
Probe voltage, RMS
U
~
Alternating voltage component
U
=
Direct voltage component
PE - P1
(with mains)
U
Probe voltage, RMS
U
~
Alternating voltage component
U
=
Direct voltage component
U
Measuring
Parameter
Meaning
Meas. Type
Suitable for
DUT Connection via
PE-P1
Measurement of voltages with
reference to PE, test socket re-
mains voltage-free
Permanent connection
PE-P1 (with mains)
Measurement of voltages with
reference to PE, line voltage is
applied to the test socket
Test socket
Polarity – only for PE-P1 (with mains)
Normal / reversed
Selection of polarity for mains voltage to the test socket
Содержание SECULIFE ST BASE 25
Страница 111: ...Gossen Metrawatt GmbH 111 ...