GOSSEN-METRAWATT GMBH
24
Differential controller
See page 28 for
Parameters
– The actual value difference = 1st actual value – 2nd actual value is controlled to the set differential set point.
– The differential set point can be set on the range
±
range span.
– Limit monitoring is referred to the difference in actual values and not the two actual values.
– If an attempt is made to adjust the differential set point on the operating level, display mode 1st actual value / 2nd actual value, by means of the
potentiometer,
no
is briefly shown on the lower display.
Slave controller
See page 28 for
Parameters
– The external set point applied to the 2nd signal input replaces the internal set point.
– The set point ramp function (see page 36) is maintained.
– When switching over to the second set point by means of a binary input, the controller becomes a
fixed value controller
with the set point
SP 2
.
– The lower and upper limit of the external set point is scaled by means of the parameters
rn L
and
rn H
(2nd standard signal for B4 and B5).
– The parameters
SP L
and
SP H
limit the external set point for control and display.
– If an attempt is made to adjust the set point on the operating level, display mode actual value / set point, by means of the potentiometer,
no
is briefly
shown on the lower display.