GMC-I Messtechnik GmbH
19
10.2.3 RISO – Probe-PE
Insulation Resistance Measurement, Probe
to Conductive Parts (which are connected to
PE)
For PC I Devices with SELV Parts
➭
Contact the SELV parts.
Insulation resistance of the SELV parts to PE
is measured with a test voltage of 250 V.
Connection
➭
Insert the test object’s mains plug into
the test socket.
➭
Connect the probe to the red probe
socket.
Procedure
➭
Press the
R
ISO
key.
➭
Select the “
Probe-PE
” menu with the help
of the
scroll keys.
➭
Start the measurement by pressing the
OK
key.
➭
Contact the low-voltage output terminals
with the probe.
➭
If the respective measured value is larger
than the specified limit value,
OK
appears
at the display and the
LED lights up
green.
➭
You can repeat the measurement by
pressing the
scroll key (reset), after
which the LED
goes out briefly until
the next measurement is performed.