Response Time
(after manual range selection)
Reference conditions
Ambient temperature 23
°
C+ 2 K
Relative humidity
45% ... 55 % RH
Frequency of
measured quantity
50 or 60 Hz
±
2%
Waveform of the
measured quantity
sinusoidal
Battery voltage
3 V
±
0.1 V
Environmental conditions
Functional
temperature range
–10
°
C ... 50
°
C
Storage
temperature range
–25
°
C ... +70
°
C (without batteries)
Relative humidity
45 ... 75 %
Altitude
up to 2000 m
Mechanical configuration
Protection for the multimeterIP50
Pollution degree
2
Connection sockets
IP20 according to EN 60529
Measured Quantity/
Measuring Range
Response Time
Transient response for
step function of the
measured quantity
Analog
Display
Digital Display
V ,
V
,
°
C
0.1 s
1 s
from 0 to 80%
of the upper range limit
A ,
A
0.1 s
1 s
from 0 to 50%
of the upper range limit
660
Ω
... 6.6 M
Ω
0.1 s
1 s
66 M
Ω
0.2 s
2 s
0.1 s
1 s
from 0 to 80%
of the upper range limit
6.6 nF ... 66 μF
0.7 s
max. 1 s
660 μF ... 6.6 mF
1.4 s
max. 3 s
66 mF
7.0 s
max. 15 s
660 Hz, 6.6 kHz
2.0 s
max. 2 s
66 kHz, 660 kHz, 1 MHz
0.5 s
max. 1 s
% (
≥
10 Hz)
0.7 s
max. 2.5 s