31
Gossen Metrawatt GmbH
V/Hz,
, Temperature,
and A/Hz Measurements
7.2 Resistance Measurement,
➭
Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭
Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to chapter 7.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭
Set the rotary switch to “
”.
➭
Connect the device under test as shown.
Note
Use short or shielded measurement cables in the case of
high-impedance resistance.
Improving Accuracy by means of Zero Balancing
Cable resistance and contact resistance can be eliminated in all
measuring ranges by means of zero balancing (see chapter 6.2).
COM
V
Temp
R
x
R
x
0 V !
!
000
.
00
k
600
60 M
Measuring Ranges: