Clamp Multimeter METRACLIP 86
English
•
In DC: depends on the remanence. Can be corrected by the "DC
zero" function of the HOLD key
•
In AC: <150mA
Specific characteristics in MAX/MIN mode
in current (from 10Hz to 1kHz in AC
and AC+DC, and from 0.30 A):
•
Uncertainties: add ± (1% R) to the values in the tables
above.
•
Capture time of the extrema: approximately 100ms.
Specific characteristics in PEAK mode
in current (from 10Hz to 400 Hz in AC
and AC+DC):
•
Uncertainties: add ± (1.5% R+0.5A) to the values in the tables above.
•
PEAK capture time: 1ms min. to 1.5ms max.
4.2.7
True-Inrush measurement
Measurement range
6 A to 600 A AC
6 A to 900 A DC
Specified measurement range
0 to 100% of the measurement range
Uncertainties
± (5% R + 5 pt)
Resolution
1 A
Specific characteristics in PEAK mode
in True-Inrush (from 10Hz to 400 Hz in
AC):
•
Uncertainties: add ± (1.5% R+0.5A) to the values in the tables above.
•
PEAK capture time: 1ms min. to 1.5ms max.
4.2.8
Continuity measurement
Measurement range
0.0Ω to 599.9
Ω
Open-circuit voltage
≤
3,6 V
Measurement current
550 µA
Uncertainties
± (1% R +5 pt)
Buzzer triggering threshold
Adjustable from 1Ω to 599
Ω
(40
Ω
is the default)
4.2.9
Resistance measurement
Measurement range (1)
0.0
Ω to
599.9
Ω
60
0 Ω to
5999
Ω
6.00 k
Ω to
59.99 k
Ω
Specified measurement
range
1 to 100% of the
measurement
range
0 to 100% of the measurement
range
Uncertainties
± (1% R +5 pt)
38