Clamp Multimeter METRACLIP 86
English
The measured value is displayed on the screen.
3.9.2
DC or AC+DC measurement
To measure the DC or AC+DC current, if the display unit does not indicate "0", first
correct the DC zero as follows:
Step 1 : to correct the DC zero
Important :
The clamp must not be closed on the conductor during the DC zero
correction. Hold the clamp in the same position during the whole procedure so that
the correction value will be exact.
Press the
key until the device emits a double beep and displays a value near
"0". The correction value is stored until the clamp is powered down.
Remark :
the correction is effected only if the value displayed is < ± 6 A, otherwise
the value displayed blinks and is not stored. The clamp must be recalibrated (see
§
5.3
)
Step 2 : to make a measurement
1.
The switch is set to
. Select DC or AC+DC by pressing the yellow
key until the desired choice is reached.
2.
Apply the clamp to only the conductor concerned.
26