Clamp Multimeter METRACLIP 86
English
2. To change the threshold, press the
key. The value flashes: each
press on the
key displays the next value. To record the chosen
threshold, apply a long press (>2s) on the
key. A confirmation beep is
emitted.
To exit from the programming mode, turn the switch to another setting. The chosen
threshold is stored (emission of a double beep).
Note: The starting current measurement triggering threshold is fixed at 1% of the
least sensitive range. This threshold is not adjustable.
3.4.4
Default configuration
To reset the clamp to its default parameters (factory configuration):
In the OFF position, hold the
key down while turning the switch to
,
until the "full screen" display ends and a beep is emitted, to enter the
configuration mode. The "rSt" symbol is displayed.
After 2 s, the clamp emits a double beep, then all of the symbols of the
screen are displayed until the
key is released. The default parameters
are then restored:
Continuity detection threshold =40Ω
True Inrush triggering threshold =10%
3.5
VOLTAGE MEASUREMENT (V)
To measure a voltage, proceed as follows :
1. Set the switch to
;
2. Connect the black lead to the
COM
terminal and the red lead to "
+
".
3. Place the test probes or the crocodile clips on the terminals of the circuit
to be measured. The device selects AC or DC automatically according to
which measured value is larger. The AC or DC symbol lights in blinking
mode.
To select AC, DC or AC+DC manually, press the yellow key to reach the desired
choice. The symbol corresponding to the choice made then lights in fixed mode.
22