INPut:PROTection:CLEar
Clears the protection status (OC, OV, OP, OT, RV) of the load.
Command Syntax: INPut:PROTection:CLEar
Parameter: None
Example: INP:PROT:CLE
Query Syntax: None
Related Commands: INP
INPut:LATCh:VOLTage[:LEVEL]
Specifies the Von voltage point. If the load is power on, it will be
enabled when the external input voltage reaches or exceeds the
Von point.
Command Syntax: INPut:LATCh:VOLTage[:LEVEL] <NRf+>
Parameter: Digits |MIN|MAX
Unit: V|mV
Example: INP:LATC:VOLT 2
Query Syntax:INPut:LATCh:VOLTage[:LEVEL]?
Parameter: None |MIN|MAX
Example: INP:LATC:VOLT?; INP:LATC:VOLT?MIN; INP:LATC:VOLT?MAX
Returned Parameter: <NR3>
Related Commands: INP:LATC
INPut:LATCh[:STATe]
Enalbes/Disables the Von Latch of the load. If the actual voltage is
less than Von point after the load has been enabled, there are two
circumstances for the load at this time: if the Von Latch is
enabled, the load continues operating; if the Von Latch is dis-
abled, the load is turned off.
Command Syntax: INPut:LATCh[:STATe] <bool>
Parameter: ON/OFF
Example: INP:LATC ON
Query Syntax: INPut:LATCh[:STATe]?
Parameter: None
Example: INP:LATC?
Returned Parameter: <NRI> Value:0(OFF) 1(ON)
Related Commands: INP:LATC:VOLT
INPut:LIMit[:CV]:CURRent
Sets the maximum current limit when the load is in the CV mode.
Command Syntax: INPut:LIMit[:CV]:CURRent <NRf+>,
Parameter: Digits |MIN|MAX
Unit: A|mA
Example: INP:LIM:CURR 20
Query Syntax: INPut:LIMit[:CV]:CURRent?
Parameter: None |MIN|MAX
Example: INP:LIM:CURR? INP:LIM:CURR? MIN
INP:LIM:CURR? MAX
Returned Parameter: <NR3>
Related Commands: CURRent
2.2.7
TRANsient Subsystem
[SOURce:]TRANsient [:STATe]
Enables/Disables the transient operation.
Command Syntax: TRANsient[:STATe] <bool>
Parameter: ON|OFF
Example: TRAN ON
Query Syntax: TRAN[:STATe]?
Returned Parameter: <NR1> Value:0 for OFF, 1for ON
TRANsient:MODE
Sets the transient operating mode. There are three transient oper-
ating modes: Continuous, Pulsed, and Toggled.
Command Syntax: TRANsient:MODE <aard>
Parameter: CONTinuous|PULSe|TOGGle
Example: TRAN:MODE PULS
Query Syntax: TRANsient:MODE?
Returned Parameter: <aard>CONT, PULS, or TOGG
Related Commands: TRIGger
TRANsient:HTIMe
Sets transient high level time.
Command Syntax: TRANsient:HTIMe <NRf+>
Parameter: Digits |MIN|MAX
Unit: mS|uS
Example: TRANsient:HTIMe 500ms
Query Syntax: TRANsient:HTIMe?
Parameter: None |MIN|MAX
Example: TRAN:HTIMe? TRAN:HTIMe? MIN TRAN:HTIMe?
MAX
Returned Parameter: <NR3>
Related Commands: TRAN:LTIM
TRANsient:LTIMe
Sets transient low level time.
Command Syntax: TRANsient:LTIMe <NRf+>
Parameter: Digits |MIN|MAX
Unit: mS|uS
Example: TRANsient:LTIMe 500ms
Query Syntax: TRANsient:LTIMe?
Parameter: None |MIN|MAX
Example: TRAN:LTIMe? TRAN:LTIMe? MIN TRAN:LTIMe? MAX
Returned Parameter: <NR3>
Related Commands: TRAN:HTIM
TRANsient:RTIMe
Sets the time for transient rising edge.
Command Syntax: TRANsient:RTIMe <NRf+>
Parameter: Digits |MIN|MAX
Unit: mS|uS
Example: TRANsient:RTIMe 500ms
Query Syntax: TRANsient:RTIMe?
Parameter: None |MIN|MAX
Example: TRAN:RTIMe? TRAN:RTIMe? MIN TRAN:RTIMe? MAX
Returned Parameter: <NR3>
Related Commands: TRAN:FTIM
TRANsient:FTIMe
Sets the time for transient falling edge.
Command Syntax: TRANsient:FTIMe <NRf+>
Parameter: Digits |MIN|MAX
Unit: mS|uS
Example: TRANsient:FTIMe 200ms
Query Syntax: TRANsient:FTIMe?
Parameter: None |MIN|MAX
Example: TRAN:FTIMe? TRAN:FTIMe? MIN TRAN:FTIMe? MAX
Returned Parameter: <NR3>
Related Commands: TRAN:RTIM