RESistance Subsystem
[SOURce:]RESistance[:LEVeI][:IMMediate][:AMPLitude]
Specifies the immediate resistance level for the load. RES? MIN
and RES? MAX return the maximum and minimum resistance
level for the present range. If the set level exceeds the range, an
error will be generated.
Command Syntax: [SOURce:]RESistance[:LEVeI][:IMMediate]
[:AMPLitude]
<NRf+>
Parameter: Digits |MIN|MAX
Unit: OHM|mOHM|kOHM
Example: RES 10 OHM
Query Syntax: [SOURce:]RESistance[:LEVeI][:IMMediate]?
Parameter: None |MIN|MAX
Example: RES? RES? MIN RES? MAX
Returned Parameter: <NR3>
Related Commands: CURRent, VOLTage
[SOURce:]RESistance[:LEVeI]:TRIGgered[:AMPLitude]
Specifies the resistance trigger level for the load. It shares the
same range with the immediate resistance level. The resistance
trigger level becomes the immediate resistance level only when
the trigger signal has been received.
Command Syntax: [SOURce:]RESistance[:LEVel]:TRIGgered
[:AMPLitude] <NRf+>
Parameter: Digits |MIN|MAX
Unit: OHM|mOHM|kOHM
Example: RES: TRIG 3 OHM
Query Syntax: [SOURce:]RESistance[:LEVel]:TRIGgered[:AMPLitude]?
Parameter: None |MIN|MAX
Example: RES:TRIG? RES:TRIG? MIN RES:TRIG? MAX
Returned Parameter: <NR3>
Related Commands: INIT, INIT:CONT
[SOURce:]RESistance:[LEVel]:LOW
Specifies the transient resistance low level, which shares the
same range with the immediate resistance level.
Command Syntax: [SOURce:]RESistance:[LEVel]:LOW <NRf+>
Parameter: Digits |MIN|MAX
Unit: OHM
Example: RES:LOW 3 OHM
Query Syntax: [SOURce:]RESistance:LOW?
Parameter: None |MIN|MAX
Example: RES:LOW? RES:LOW? MIN RES:LOW? MAX
Returned Parameter: <NR3>
Related Commands: RESistance:HIGH, TRAN:HTIMe
[SOURce:]RESistance:[LEVel]:HIGH
Specifies the transient voltage high level, which shares the same
range with the immediate resistance level.
Command Syntax: [SOURce:]RESistance:[LEVel]:HIGH <NRf+>
Parameter: Digits |MIN|MAX
Unit: OHM
Example: RES:HIGH 3 OHM
Query Syntax: [SOURce:]RESistance[:LEVEL]:HIGH?
Parameter: None |MIN|MAX
Example: RES:HIGH? RES:HIGH? MIN RES:HIGH? MAX
Returned Parameter: <NR3>
Related Commands: RESistance:HIGH, TRAN:HTIMe
2.2.5
POWer Subsystem
[SOURce:]POWer[:LEVeI] [:IMMediate][:AMPLitude]
Specifies the power level. POW? MIN and POW? MAX return the
maximum and minimum power level for the present range. If the
set level exceeds the range, an error will be generated.
Command Syntax: [SOURce:]POWer[:LEVeI][:IMMediate]
[:AMPLitude]<NRf+>
Parameter: Digits |MIN|MAX
Unit: W|mW
Example: POW 10W
Query Syntax: [SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude]?
Parameter: None |MIN|MAX
Example: POW? POW? MIN POW? MAX
Returned Parameter: <NR3>
Related Commands: POWer:TRIGgered
[SOURce:]POWer[:LEVeI]:TRIGgered[:AMPLitude
]
Specifies the power trigger level. It shares the same range with
the power level. The power trigger level becomes the power level
only when the trigger signal has been received.
Command Syntax: [SOURce:]POWer[:LE'Vel]:TRIGgered
[:AMPLitude]<NRf+>
Parameter: Digits |MIN|MAX
Unit: W|mW
Example: POW:TRIG 10w
Query Syntax: [SOURce:]POWer[:LEVel]:TRIGgered[:AMPLitude]?
Parameter: None |MIN|MAX
Example: POW:TRIG? POW:TRIG? MIN POW:TRIG? MAX
Returned Parameter: <NR3>
Related Commands: INIT, INIT:CONT
2.2.6
INPut Subsystem
INPut[:STATe]
Enables/Disables the load input.
Command Syntax: INPut[:STATe] <bool>
Parameter: ON |OFF
Example: INP ON
Query Syntax: INPut[:STATe]?
Parameter: None
Example: INP?
Returned Parameter: <NRI> Value:0(OFF) 1(ON)
Related Commands: INP:SHOR
INPut:SHORt[:STATe]
Enables/Disables the short circuit of the load.
Command Syntax: INPut:SHORt[:STATe] <bool>
Parameter: ON |OFF
Example: INP:SHOR ON
Query Syntax: INPut:SHORt[:STATe]?
Parameter: None
Example: INP:SHOR?
Returned Parameter: <NR1> Value:0(OFF) 1(ON)
Related Commands: INP[:STAT]