4.14.2 Short Circuit Operation
Example:
Short circuit operation in CV mode. The detailed operations are shown below:
The detailed operations are shown below:
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT
ON
Sets the load to short circuit
MODE
CV
Sets the mode
INPUT
ON
Turns on the load
4.14.3 Von Point/Von Latch
Example:
Set the Von Latch to Off; set Von Point to 1 V; turn on the load; This example implements the automatic turning on/off of the
load‘s input, which simplifies the test operations greatly. The detailed operations are shown below:
For the above steps, the corresponding SCPI commands should be:
INP:VOLT:ON:LATCH
OFF
Disables the Von Latch
INP:VOLT:ON
1
Sets the Von voltage to 1 V
INPUT
ON
Turns on the load
Procedure
Description
Display
1
Press the
key to enter into the main menu.
Load Default
:
Yes *No
2
Use the
▼
key to select the menu item “Short”; use the knob or
◄
key to select the parameter “On”.
Press the
Enter
key to confirm.
Short
On *Off
3
Press the
key to exit the main menu.
4
Press the
key to enter into the mode selection and parameter setting menu; use
▲ ▼
keys to select CV mode.
Press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
0.000V sCV
OFF
5
Press the
key to turn on the load.
0.000V 0.000A
0.000V sCV
ON
Procedure
Description
Display
1
Press the
key to enter into the main menu.
Load Default
:
Yes *No
2
Use the
▼
key to select the menu item “Von Latch”; use the knob or
►
key to select the parameter “Off”.
Press the
Enter
key to confirm.
Von Latch
:
On *Off
3
Use the
▼
key to select the menu item “Von Point”; use Entry keys or the knob together with
▲ ▼
keys to input the voltage value.
Press the
Enter
key to confirm.
Von Point
:
1.000V
4
Press the
key to exit the main menu.
0.000V 0.000A
0.000A CCH
OFF
5
Press the
key to turn on the load.
0.000V 0.000A
0.000A CCH
ON