Gossen Metrawatt GmbH
2.4.3
Toggled Transient Mode
The trigger function is required for toggled transient operation. As long as no trigger occurs, the load remains at transient low level.
After a trigger has been received, a toggle operation will be executed, and high transient level will be reached after the duration of rising
edge. The load will keep this level until a new trigger is received. Once this happens, the load will return to transient low level after the
duration of falling edge. The load will switch between the levels each time a trigger is received.
The associated parameters such as
– low level (LevelL),
– high level (LevelH),
– time for rising edge (TimeR),
– time for falling edget (TimeF) and
– toggled transient mode (toggle)
can be set through transient operation menu or via remote command
– (CURRent:LLEVel <NRf+>,
– CURRent:HLEVel <NRf+>,
– VOLTage:LLEVel <NRf+>, VOLTage:HLEVel <NRf+>,
– RESistance:LLEVel <NRf+>,
– RESistance:HLEVel <NRf+>,
– TRANsient:RTIMe <NRf+>,
– TRANsient:FTIMe <NRf+>,
– TRANsient:MODE TOGGle).
Low level time (TimeL) and high level time (TimeH) have no influence on toggled transient mode.
The trigger can be an external trigger signal received via the TRIG input on the front panel, pressing the
key and
key,
or via the
GPIB GET
function, the common command
*TRG
or the subsystem command
TRIG.
For example: assume that the CCH range is active, and the input is in OFF status, then the transient parameters should be set as fol-
lows:
Press the
key to enter into transient operation.
Press the
key to open transient operation menu.
Set the following parameters in transient operation menu:
Then press the
key to turn on the input.
Alternatively, set the parameters via remote command:
Get the trigger by receiving an external trigger signal. Figure 2.8 shows the current waveform of the load before it is triggered and after
it has been triggered respectively: the electronic load starts its operation at the transient low level (5A) when the input is turned on. For
the first trigger, the load current reaches and remains at the high level (10A) after 100 μs duration of rising edge. When the second trig-
ger is received, the load current will reach and remain at the low level (5A) after 200 μs duration of falling edge. Each trigger leads to
one toggle operation.
LevelL
5.000 A
LevelH
10.000 A
TimeR
0.10 ms
TimeF
0.20 ms
Mode
Togg
SCPI Command
Description
TRIG:SOUR EXT
Selects the external trigger input
TRAN ON
Enables transient operation
TRAN:LLEV 5
Sets transient current low level to 5A
TRAN:HLEV 10
Sets transient current high level to 10A
TRAN:RTIM 100us
Sets the time for transient rising edge to 100 μs
TRAN:FTIM 200us
Sets the time for transient falling edge to 200 μs
TRAN:MODE TOGG
Selects toggled operating mode
INPUT ON
Turns on the input