Goodson MCB-90 Скачать руководство пользователя страница 1

MCB-90, DMBG-50160  

90º Dial Bore Gauge

Instructions

Precautions

•  Do NOT disassemble the instrument
•  Do NOT bump any part of the instrument
•  After use, clean the interchangeable rod/washer, etc., apply a coat of anti-corrosive oil 

and store them in their original containers

•  Use only the supplied parts and accessories. Do NOT use then with other 

instruments

Install the Dial Indicator

Install the dial indicator by inserting its spindle into the holder so that the 

indicator reads at least 0.3mm. Secure the dial indicator with the clamp 

screw (Fig. 1).

Attach The Limit Hand 

To install the limit hand (where provided), hook 

it on the lower bezel and push it until it clicks 

into place (Fig. 2). To remove it, force it open.

Set The Measuring Dimension

To set the required measuring dimension, select the appropriate interchangeable 

rod, washers and sub-anvils and install them on the main unit as shown in Figure 3. Always use a 

minimum number of rods, washers and sub-anvils to achieve the size you require.

Zero Adjustments With A Setting Ring or Master Gauge

The setup for the bore gauge is the same as that for an actual measurement. 
Diameter (e) is maximized when intersection (Fig. 4A) is perpendicular to the axis of the objective 

cylinder. In contrast, diameter (e) is minimized at intersection (Fig. 

4B) which includes the cylinder axis so the bore gauge takes 

the maximum reading. When using the bore gauge, adjust the 

movement of the contact point along (a) on the intersection (Fig. 4A) 

and along (e) on intersection (Fig 4B).
To adjust the datum point to diameter (e), search for a position 

where the bore gauge reading is minimized by 

rocking the bore gauge back and forth (Fig. 5). 

Last, adjust the datum point by turning the bezel 

unit until it is aligned with the indicator hand, 

representing the obtained diameter (Fig. 6). 

Zero Setting With An Outside Micrometer

Mount the micrometer firmly on the stand, then set the opening 

to an appropriate size. Place the bore gauge into the micrometer 

opening and find the position where it indicates the minimum 

value. Adjust the datum point by rotating the bezel. Skill is 

required to adjust the datum point using the micrometer as it is not automatically set to proper 

position by the guide. NOTE: Hold the micrometer vertically so 

that the micrometer head is facing downward. Do NOT clamp the 

micrometer while performing this adjustment (Fig. 7).

© Goodson Tools & Supplies 

April 2016

Figure 1

Bore Gauge

Clamp Screw

Figure 2

Figure 5

Figura 4

Figure 6

Sub-anvil

Washer

Figure 3

Figure 7

Отзывы: