10 Servicing of the OSMOMAT auto
■
Ribbon Change
The ink ribbon cassette must be replaced when the printout no longer has sufficient contrast.
1. Tear off any paper protruding from the slot before replacing the ribbon.
2. Press lightly on the right side of the ink ribbon cassette.
The cassette is released from the printer.
3. Take out a new ribbon and
use your index finger to twist the knob clockwise
and tighten the ribbon.
4. Insert the ribbon in the printer. The paper must be
between the ribbon and the ribbon cassette.
Fig. 8 Replacing the ribbon
Tighten the ribbon by twisting the knob again.
5. Press and hold the
Paper Feed
button until approx. 2 cm of the printer paper is protruding
from the housing and then close the door.
10.7 Replacing the Cleaning Paper
A red marker on the cleaning paper indicates the end of the paper roll during operation.
The remaining paper is sufficient for two or three more charge measurements. After that, the roll must
be replaced.
Corrupt measurement results: A check for cleaning paper is not performed
automatically. The system also works without cleaning paper. In this case, the sensor is
not cleaned and the samples mix with sample residues from previous sample
measurements.
1. Cut the end of the paper roll at an angle.
40 | Servicing of the OSMOMAT auto