Calibration Using Calibration Panel CP 40
Sensor Calibration
Page 32 (40)
at
os
_cs
_r
ev
01_en
_r
ev
-c
2012
-09
-03
10.3.4
Calibration Process
After that the software will guide you through the individual calibration
steps. In the first step, you are asked to place the sensor and the cali-
bration object in the center of the measuring volume.
Enable the
Automatic exposure time
.
In exceptional cases it might happen that the function
Automatic ex-
posure time
cannot work correctly. In this case, disable the function
and use the exposure time wheel to adjust the optimum exposure time
for each calibration image.
Exposure time adjustment for calibration
An optimum exposure time is given when the red overexposed areas
are just no longer visible.
In the subsequent steps you are then asked to place the calibration
object in defined positions to the sensor. The software leads you to the
correct positions.
10.3.5
Calibration Result
At the end of the calibration process, the calibration result is shown.
The green icon indicates a good calibration result. The red icon
indicates an insufficient calibration result.
In case of an insufficient calibration result, repeat the calibration and
check the sensor settings (see 8.4), the correct camera lenses and
ensure that the cameras and lenses are mounted tight.
You reach the calibration results with
Acquisition
►
Sensor
►
Cali-
bration
►
Show Calibration Information
.
As of software version V7.5 and higher, the calibration is saved in the
ATOS COMPACT SCAN SENSOR.