Hardware
basicCAN 61 PLUS – User Manual
3-17
basicCAN 61 PLUS
units provide a 1Gbit Ethernet interface and a USB 2.0
interface. Both interfaces can be used for the communication of the unit
with the host PC.
In case of using the
Ethernet
interface, the device can be controlled via
the
default IP Address
192.168.1.62
,
Port 5134
, which can be changed if
required. In principle, there are two ways for this:
♦
HardwareExplorer: Select the device, under
Device
set the
required
IP Address
;
the new
IP Address
is effective after restart
♦
G API Command
G_Common_Ethernet_IpAddress_Set
;
the new
IP Address
is effective after restart
Addressing of
basicCON 61 PLUS
devices controlled via the USB
interface takes place exclusively according to their serial numbers: The
device with the least serial number is always
the device with the
number
1.
To improve clarity, we recommend to connect several
basicCAN 61 PLUS
devices (if applicable) in the order of ascending
serial numbers to the PC/ Laptop.
Electric surges can harm expensive test equipment and may lead to
unreliable test results. Electric isolation protects against electric surges
and can help to suppress dangerous electrical transients. It also
eliminates ground loops, responsible for data errors due to ground
potential differences.
A
basicCAN 61 PLUS
unit provides electric isolation between the USB/
Ethernet system and all input and output signals of the frontal
connectors. This includes the CAN, LIN, K-Line and FlexRay
communication interfaces as well as digital and analog IOs.
3.5.9
Addressing
3.5.10