I S A – D a t a T r a n s m i s s i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 41 / 76
www.go-sys.de
SQI (Spectral Quality Index)
The SQI is a degree of the statistical reliability of measurement values and is calculated continuously for
each parameter. Precondition is the creation of a corresponding calibration file in the xml format.
see
Appendix D – SQI (Spectral Quality Index)
Imports ISA Plus calibration data in the txt-format and in the xml-format.
Transmits the settings to the spectrometer unit.
10 Transmitting the Data to the PC
There are two ways of transmitting the spectrometer data from the BlueBox to a PC.
•
Transmitting with
BlueBox SQL
Software
Here the
whole data
of the BlueBox are transmitted to the PC.
•
Transmitting with
ISA Data
Here only spectra data is transmitted.
Both ways save the data in a SQL data base. The memory location of the SQL data base was determined at the
installation of the SQL server.
Opens the window of the quality settings, only visible if the SQI is calculated.
In this window the calculation values of the SQI are
displayed.
Only the value for
Max h
is of interest for the opera-
tion, h is the
value of the SQI
. If the
Max h
is ex-
ceeded, the measurement value is displayed in
square brackets in the sensor setup window and in
the display of the BlueBox.
In addition, the sensor icon is displayed as a warning
sign
in the AMS start window.