5-7
VIEW / PRINT TEST RESULTS
Test Result Files:
The test file names in the list, (e.g. PS_200_200207
_01_9_2011_1450.xml) identifies serial number of the
instrument (200207), the date tested in day, month, year
(e.g. 01_09_2011) and the time testing was completed (e.g.
1450hrs).
Each ‘.xml’ file can be opened individually to display
a comprehensive list of test results obtained from that
particular instrument.
To open a test file, position cursor on the file name then double
click mouse button to select.
The following example, Fig. 5-11, illustrates information
available, including:
• Instrument Type
(e.g. GMI PS200)
• Test Type
(e.g. Calibration)
•
Instr. Serial No
(e.g. 200255)
• New Cal Due Date (if Test registered as PASS)
• Station Serial No.
(e.g. 000101)
• Date Tested
(e.g. 06 September 2011)
• Time Tested
(e.g. 2:30:04 PM)
• Overall Result
(e.g. PASS)
Followed by a listing of gas input / target / Lo Limit / Hi Limit
/ result for each gas range.
Finally, an audible / visual alarm test result is displayed.
Содержание PS200 ABC
Страница 1: ...Personal Surveyor 200 Auto Bump Calibration Station Handbook Gas Measurement Instruments Ltd...
Страница 6: ...iv AUTO BUMP CAL STATION HANDBOOK...
Страница 10: ...viii AUTO BUMP CAL STATION HANDBOOK...
Страница 16: ...1 6 AUTO BUMP CAL STATION HANDBOOK...
Страница 26: ...3 2 AUTO BUMP CAL STATION HANDBOOK Fig 3 2 Windows Vista Fig 3 3 Windows 7...
Страница 54: ...4 14 AUTO BUMP CAL STATION HANDBOOK...
Страница 62: ...5 8 AUTO BUMP CAL STATION HANDBOOK EXAMPLE Fig 5 11...
Страница 69: ...5 15 VIEW PRINT TEST RESULTS Fig 5 17 Bump Results...
Страница 70: ...5 16 AUTO BUMP CAL STATION HANDBOOK Fig 5 18 Calibration Certificate...
Страница 79: ......