19
‘SAMPLE FAULT’
This indicates a problem with the instrument’s flow due
to the sample path being blocked, water ingress, a blocked
filter or pump failure.
In normal mode the pump stops automatically. The sample
line, filters etc. should be checked for water ingress or
blockage and Switch One should then be used to restart
the pump.
‘CHECK ZERO’
This indicates that there may have been a zero shift due
to the presence of gas. Switch off the instrument and
switch on again in fresh air.
‘ZERO FAULT’
This indicates that the zero is outwith its calibration limits.
Switch the instrument off and then on again in fresh air.
If the fault does not clear, leave the instrument on for a
period of 30 minutes then switch off and on again. If the
fault still does not clear, return the instrument for
servicing.
‘BAT’
This indicates that the batteries will soon require
replacement. At this point there will be approximately 60
minutes operation left in a set of alkaline batteries,
although this figure will vary depending on battery
manufacturer, temperature conditions, usage etc. With
rechargeable batteries the ‘BAT’ flag indicates
approximately 30 minutes operation left.
OPERATION
Содержание PPM Gasurveyor 500
Страница 1: ...PPM Gasurveyor 500 User Handbook...
Страница 6: ...PPM GASURVEYOR 500 USER HANDBOOK iv...
Страница 22: ...PPM GASURVEYOR 500 USER HANDBOOK 12...
Страница 48: ...PPM GASURVEYOR 500 USER HANDBOOK 38...
Страница 52: ...PPM GASURVEYOR 500 USER HANDBOOK A 4...
Страница 62: ...PPM GASURVEYOR 500 USER HANDBOOK B 10...
Страница 84: ...C 22 GAS MEASUREMENT INSTRUMENTS LTD...
Страница 88: ...PPM GASURVEYOR 500 USER HANDBOOK IV...
Страница 89: ......