![Giropes 261418 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/giropes/261418/261418_manual_2229285018.webp)
18
EN
EN
USER MANUAL
BAP
4
TWB
V3
Tare weight
W/C/%
6
5
GWA
V4
Gross weight (with “.”)
W/C/%
7
6
GWB
V5
Gross weight
W/C/%
6
7
QUA
V6
Quantity/Percentage
(.)
C/%
7
8
QUB
V7
Quantity/Percentage
C/%
6
9
UWA
V8
Unit weight (with “.”)
C
7
10
UWB
V9
Unit weight
C
6
11
UNT
V10
Weighing Unit
W/C/%
3
7. CALIBRATION PROCEDURE
Switch on the balance, and press the CAL switch during the self-test.
CAL SWITCH
When the self-test procedure finished, the display will show as below:
CAL
When it shows “CAL”, press
key to start calibration procedure.
0.00
g
•
Touch the pan lightly by hand, and several seconds later, once it take stability again, the display shows the
weight value of first calibration mass (1/3 of full capacity). Place the displayed weigh on the pan.
100.00
g
•
Several seconds later, once it take stability, the display shows the weight value of second calibration mass
(2/3 of full capacity) Place the displayed weigh on the pan.