background image

MISURATORE DI PRESSIONE GIMA BLUETOOTH

GIMA BLUETOOTH B.P.MONITOR 

TENSIOMÈTRE BLUETOOTH GIMA

MEDIDOR DE PRESIÓN DEL BLUETOOTH DEL GIMA

M32916-M-Rev

.0-07.20

PROFESSIONAL MEDICAL PRODUCTS

Guangdong Transtek Medical Electronics Co., Ltd.

Zone A, No.105, Dongli Road, 

Torch Development District,

Zhongshan, 528437, Guangdong, China

Made in China

32916 / LS808-BS

MDSS - Medical Device Safety Service GmbH

Schiffgraben 41, 30175 Hannover, Germany

0123

Gima S.p.A. 

Via Marconi, 1 - 20060 Gessate (MI) Italy

[email protected] - [email protected]

www.gimaitaly.com

Содержание LS808-BS

Страница 1: ...MEDICAL PRODUCTS Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District Zhongshan 528437 Guangdong China Made in China 32916 LS808 BS MDSS Medical Device S...

Страница 2: ...DATA MANAGEMENT 44 Recall the Records 44 Delete the Records 46 INFORMATION FOR USER 47 Tips for Measurement 47 Maintenance 48 ABOUT BLOOD PRESSURE 49 What are systolic pressure and diastolic pressure...

Страница 3: ...ntended for adult indoor use only Contraindications 1 The device should not be used by any person who may be suspected of or is pregnant 2 The device is not suitable for use on patients with implanted...

Страница 4: ...r blood pressure When the device is used to measure patients who have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation the best result may occur with deviation P...

Страница 5: ...use in the presence of a flamma ble anesthetic mixture with air of with oxygen or nitrous oxide Warning No servicing maintenance while the ME equipment is in use The patient is an intended operator Th...

Страница 6: ...ort to Gima if any unexpected operation or events occur Keep the unit out of reach of infants young children or pets to avoid inhalation or swallowing of small parts It is dangerous or even fatal Be c...

Страница 7: ...inks when the blue tooth is working Average value The average value of the latest three groups bood pressure value Memory Query Indicate it is in the memory mode and which group of memory it is User 1...

Страница 8: ...sure Monitor 2 AC Adaptor LS808 BS Model BLJ06L060100P V 3 User Manual 4 Cuff 22cm 42cm Type BF Applied Part Please use Gima authorized cuff The size of the actual cuff please refer to the label on th...

Страница 9: ...d for about 20 days Storge and use the blood pressure monitor at the cool dry and ventilated environment Avoid to approach to the fire and the heat source or it will cause the battery explode Only can...

Страница 10: ...Setting the Time Date and Unit To ensure the stored measurement result has correct time record please set time and unit before device is used Before use switch the button to the ON side to turn on th...

Страница 11: ...39 3 Press User 1 button again to confirm YEAR Then the numeral representing MONTH blinks 4 Repeat step 2 and 3 to confirm MONTH and DAY 5 Repeat step 2 and 3 to confirm HOUR and MINUTE ENGLISH...

Страница 12: ...ure Monitor with Your Device 1 Turn on Bluetooth and the app Make sure both are ON when pair up is proceeding 2 In the search device interface please select the device to pair up 3 When the monitor is...

Страница 13: ...will shut off after Pair up process is complete Bluetooth Module No LS51802 RF Frequency Range 2402 MHz to 2480 MHz Output Power Range 0 dBm Supply Voltage 1 8 3 6 V Transmitting Distance 10 meters Li...

Страница 14: ...n in your arm use the other one 2 Roll or push up your sleeve to expose the skin 3 Apply the cuff to your arm with your palm facing up 4 Position the edge of the cuff about 2cm 3cm from elbow 5 Fasten...

Страница 15: ...sure data Take User 1 for example Note Select the same user on your app and BPM to take the measurement or the measurement data won t be transmitted to the app 1 When the monitor is off press the User...

Страница 16: ...DATA MANAGEMENT Recall the Records 1 When the monitor is off please press the Query button to show the average value of the latest three records If the records are less than three groups it will displ...

Страница 17: ...is January 2nd is 8 00 CAUTION The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and...

Страница 18: ...60 is dropped from the list Delete the Records 1 Hold pressing Query button for 3 seconds when the monitor is in the memory recall mode the dEL ALL User ID will flash on the display 2 Press Query to...

Страница 19: ...taken in the following circumstances Within 1 hour after dinner or drinking Immediate measurement after tea coffee smoking Within 20 minutes after taking a bath When talking or moving your fingers In...

Страница 20: ...t in a dry place and avoid the sunshine Avoid immersing it in the water Clean it with a dry cloth in case Avoid intense shaking and collisions Avoid dusty environment and unstable temperature surround...

Страница 21: ...ysician can tell whether your blood pressure value has reached a dangerous point Irregular Heartbeat Detector An irregular heartbeat is detected when a heartbeat rhythm varies while the device is meas...

Страница 22: ...the pressure will vary more 3 Wait at least 3 minutes for another measurement Why do I get a different blood pressure at home compared to the hospital The blood pressure is different even throughout...

Страница 23: ...ust the cuff and relax for a moment and then measure again E 10 or E 11 shows The monitor detected motion talking or the pulse is too poor while measuring Relax for a moment and then measure again E 2...

Страница 24: ...ange of 700 hPa to 1060 hPa Storage transportation condition Temperature 20 C to 60 C A relative humidity range of 93 non condensing at a water vapour pressure up to 50hPa Measurement perimeter of the...

Страница 25: ...V Output 100 240V 50 60Hz 0 2Amax Input 6V 1000mA Contact Information Manufactured by Guangdong Transtek Medical Electronics Co Ltd Company Guangdong Transtek Medical Electronics Co Ltd Address Zone...

Страница 26: ...al standard Electromagnetic disturbances Requirements and tests Performance requirements EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 1 Requirements and test methods for non automated measu...

Страница 27: ...l antennas should be used no closer than 30 cm 12 inches to any part of the equip ment LS808 BS including cables specified by the manufacturer Otherwise degradation of the performance of this equipmen...

Страница 28: ...rrup tions and voltage variations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 0 UT 250 300 cycle...

Страница 29: ...0 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2...

Отзывы: