Cardiogima 1M QUICK GUIDE
Rev. Feb 2008
Cardiogima 1M QUICK GUIDE
Rev. Feb 2008
GIMA SPA
www.gimaitaly.com
GIMA SPA
www.gimaitaly.com
(1) Paper Feed
(2) Sequential Mode or Printing Memory Data
(3) Manual Mode
(4) Isochronous (Automatic) Mode
(5) Paper Speed
(6) Sensitivity of Recording
(7) Anti Muscle Tremors Filter
(8) Anti Base Line Disturbs Filter
(9) Anti Power Line Disturbs Filter
\
(10/11) Vertical Arrovs: Traces Group Selection
Modes Menu Selection
(12) Modes Menu
4
Power on/off:
To turn on/off the instrument press the main switch, placed on its back side. The
initials screen will display the battery status bar.
If the charge value is lower than 80% please plug in the charger.
To start the work press any key.
Default operational set-up:
At power on the default set-up is:
•
Traces print number: 3
•
Traces Order: Standard
•
Paper speed: 25mm/s
•
Sensitivity of recording: 5mm/mV
•
Filters: Disabled
•
Hardcopy frame length: 5 seconds
The main operational modes:
1.
Paper speed:
25mm/s at the turning on. Pushing the key
(5)
it is possible to
select the paper speed: 50mm/s, 5mm/s, 10 mm/s,25mm/s…..5mm/s……
The mode appears on the display.
2.
Sensitivity of recording:
10mm/mV by default. Pushing the key
(6)
it is possible
to select the sensitivity of the trace: 20mm/mV, 5mm/mV,…..10mm/mV….
The mode appears on the display.
3.
Filters:
disabled by default. Pushing the keys (7); (8); (9); it is possible to enable
the desired filter:
-
LP
: anti muscle tremors filter
-
HP
: anti base line disturbs filter
-
N5/N6
: anti power line disturbs filter (N5 for 50 Hz and N6 for 60 Hz)
4.
Traces Group:
the Standard modality is predefined at power on and the display
showes the I trace.
It is possible to select the desired trace pushing the keys (10) and (11)
.
With the
Standard Modality
the display will show: I; II; III; aVR; aVL; aVF;
V
1
; V
2
; V
3
; V
4
; V
5;
V
6
Changing to
Cabrera Modality
the display will show
:
aVL; I; -aVR; II; aVF; III;
V
1
; V
2
; V
3
; V
4
; V
5
; V
6
1
How to open the thermal
printer
Lift the lever, acting as indicated
by the arrow, until the cover
group is released from its locking
position.
To avoid damages to the lever do
not use excessive force