3-14
Operation
3
Configuring the Liquid Handler
Safety Contact
The instrument has provisions for connecting safety
devices that your application may require, as long
as they present a contact closure or TTL type
interface. The Safety Contact tab allows you to
specify which input contact is connected to the
safety device and what is the active state of that
device. The function of the safety contact is
equivalent to pressing the Stop button on the
instrument’s front panel.
Bearing Life
The Bearing Life tab displays the XYZ travel in
kilometers.
Содержание Quad-Z 215
Страница 1: ...Quad Z 215 Liquid Handler User s Guide LT2545 2003 Gilson Inc All rights reserved April 2003...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 15: ...1 7 Introduction 1 Technical Specifications...
Страница 16: ...1 8 Introduction 1 Technical Specifications...
Страница 17: ...2 1 2 This section takes you through the steps for setting up your Quad Z 215 Liquid Handler Installation...
Страница 72: ......
Страница 73: ...5 1 5 Troubleshooting...
Страница 82: ......