Plumbing Connections
make all connections as described in the user’s guide, except the transfer tubing connection,
as this is different with the PLC Purification System. refer to the
GX-241 Liquid Handler
User’s Guide
and the
VERITY® 4X20 Syringe Pumps User’s Guide
. Complete instructions for
each step are included in the ‘installation’ chapters.
PLC Purification System with GX-241 Liquid Handler Plumbing Connections
Connection
tubing
fittings
A
from
reservoir
Solvent inlet tubing assembly
• 3 mm (oD), 2 mm (iD)
(Pn 499484021)
20 µL filter
(Pn 4957226)
TO
Syringe pump inlet
PEEK
B
from
Syringe pump outlet
transfer tubing
• 5.5 mL for 5 mL syringe
(Pn 499671112)
• 10.5 mL for 10 mL syringe
(Pn 499474103)
• 30 mL for 25 mL syringe
(Pn 499483602)
PeeK
to
Port 2
automated sample loop
injection valve
CONTINUED ON PAGE 97
A
B
C
D
E
F
G
PLC Purification System
GX-241 Liquid Handler
VERITY® 4020
Single Syringe
Pump
Reservoir
Embedded
Detector
Embedded Pump
System
1
2
3
4
5
6
H
J
K
L
Figure 72
Plumbing Diagram for the PLC Purification System with GX-241 Liquid Handler and VERITY® 4020 Single Syringe Pump
Содержание PLC 2050
Страница 1: ...PLC 2050 2250 2500 Purification Systems User s Guide...
Страница 6: ...PLC 2050 2250 2500 Purification Systems USER S GUIDE...
Страница 34: ......
Страница 50: ......
Страница 58: ......
Страница 70: ......
Страница 78: ......
Страница 88: ......
Страница 102: ......
Страница 112: ......
Страница 130: ......
Страница 131: ......
Страница 132: ...www gilson com ContactUs LT380160 01...