background image

Rev: 06/15/2015

Four-Station Proportional Caliper

HM-38B

PHONE:  800-444-1508                                                 

        P.O. Box 200, Lewis Center, Ohio  43035-0200 

FAX:      800-255-5314       

       

740-548-7298 

           E-mail: [email protected]     Website: www.globalgilson.com 

                740-548-5314  

                  

OPErAtiNg MANuAl

 

iNtrOduCtiON:

 These instructions are a general guide for operation of this 
device. The complete procedure for determination of relative 
shape characteristics of coarse aggregates using the Four-
Station Caliper is described in ASTM D4791. The user should 
carefully review the ASTM procedure before testing.

 

OPErAtiNg iNStruCtiONS:

 Individual particles of aggregate of specific sieve sizes are 
measured to determine the ratios of width to thickness and 
length to width. The applied ratios are 1:2, 1:3, 1:4, and 1:5 
according to desired specifications.

A.  For Classification of Flat Particle Testing:

 
The user should set the larger opening of the Caliper 
equal to the particle width. If the thickness of the same 
particle can be placed within the smaller opening without 
readjusting the Caliper, the particle is classified as 

Flat.

HM-38B

A. Test for Flatness

B. Test for Elongation

(Continued on back.)

ASTM D4791

B.  For Classification of Elongated Particle Testing:

The user should set the larger opening of the Caliper equal 
to the particle length. If the width of the same particle can 
be placed within the smaller opening without readjusting 
the Caliper, the particle is classified as 

Elongated.

 If the particle does not meet either classification of Flat or 
Elongated, it is classified as 

Neither Flat nor Elongated.