Series 8650B Universal Power Meters
3-4
Publication 31470-001, Rev C, November 2, 2017
3.1.3
Data Output Formats
The data output format for the standard measurement collection mode is:
±D.DDDDE±NNCRLF
±
:
Sign of the Mantissa
D.DDDD
:
Mantissa (5 digits)
E
:
Exponent (indicates that an exponent follows)
±
:
Sign of the Exponent
NN
:
Magnitude of the Exponent
CR
:
Carriage Return
LF
:
Line Feed
Data output formats for the GPIB Swift and GPIB Fast Buffered modes are expressed in the form of a signed five-
digit number with two digits to the right of the decimal and no exponents. In some cases multiple values are
sent:
One sensor Swift mode:±DDD.DD CRLF
Two sensor Swift mode:±DDD.DD,±DDD.DD CRLF
Fast Buffered mode:±DDD.DD, . . . . .±DDD.DD CRLF
3.1.4
Power-On Default Conditions
The interface wake-up state is:
•
GPIB Local Mode
•
Unaddressed, Service Request Mask Cleared
•
Status Byte Cleared
•
TR3 Free Run Trigger Mode Set
•
GT2 Group Execute Trigger Mode Set
•
Parallel Poll Data Lined Unassigned
•
Display Enabled
•
Service Request Mask Cleared
•
Event Status Register = 128
•
Event Status Mask Clear
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...