Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-141
3.19.15
Store & Recall
The instrument’s current configuration can be saved in a register for later recall.
3.19.15.1
Saving a Configuration
The commands for saving the instrument state are based on the ST function code:
Syntax:
ST [r] EN
ST is the Save function code
[r] identifies the register in which the instrument’s configuration is to be saved (and from which it
can later be retrieved). The value of [r] can be any number from 1 through 20
EN is a required terminating suffix
Example:
ST 12 EN
! SAVES THE CURRENT INSTRUMENT CONFIGURATION IN REGISTER 12
NOTE:
Register 0 contains the previous state of the instrument; to recover from an accidental preset,
recall the configuration from that register.
3.19.15.2
Retrieving a Configuration
The commands for recalling a configuration from a register are based on the RC function code:
Syntax:
RC [
r
] EN
RC is the Recall function code
[
r
] identifies the register in which the instrument’s prior configuration has been saved and from
which it is now to be retrieved. The value of [
r
] can be any number from 0 through 20
EN is a required terminating suffix
Example:
RC 3 EN
! RECALL THE CONFIGURATION THAT WAS SAVED IN REGISTER 3
Example 2:
RC 0 EN
! RECALL THE PRIOR CONFIGURATION (THIS COMMAND RECOVERS
! FROM CONFIGURATION ERRORS)
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...