Series 8650A Universal Power Meters
4-4
Manual 31470, Rev. E, April 2001
4.3.2
Calibrator Frequency Check
To measure the frequency of the calibrator:
1.
Connect the frequency counter to the calibrator output connector.
2.
Turn ON the calibrator according to the procedure given in Procedure Step 8 on the previous
page.
3.
Measure 49 to 51 MHz.
4.
Set the calibrator frequency to 1.000 GHz (with Option 12 only).
5.
Measure 999.5 MHz to 1.0005 GHz (with Option 12 only).
6.
Turn the calibrator OFF.