Series 8650A Universal Power Meters
3-40
Manual 31470, Rev. E, April 2001
Status Byte Register
Notes:
1.
The condition indicated in bits 1 through 5 must be enabled by the Service Request Mask to cause a Service Request
Condition. The mask is set with the *SRE command followed by ASCII characters. The value of the byte is determined by
summing the weight of each bit to be checked.
2.
The RQS (bit 6) is true when any of the conditions of bits 4, 5, and 7 are enabled and occur.
3.
Bits remain set until the Status Byte is cleared.
3.2.22
Min/Max Configuration and Monitoring
CALCulate:MAXimum
CALCulate:MINimum
MIN and MAX channel monitoring records the minimum and maximum variation of channel
amplitude over time. Time zero is set by setting the MIN or MAX STATe to ON. The default Preset
and turn on configuration is ON. After setting to ON, the minimum and maximum value of the
measured value on that channel will be recorded. To reset the MIN and MAX value back to the current
value, you only need to set the STATe to ON again. The STAT OFF does not need to be sent during
reset of MIN and MAX values. Individual sensor minimum and maximum power level variations can be
monitored only if that sensor is defined directly to a channel with POW 1 or POW 2. MIN and MAX
channel monitoring is only available in Normal Mode. BURSt and SWIFt modes do not perform MIN
and MAX monitoring.
CALC:MAX
Syntax
:
CALCulate<channel 1 or 2>:MAXimum:STATe
space
<ON or OFF>
Example
:
OUTPUT @Pwr_mtr;CALC2:MAX:STAT ON
! Enable channel 2 maximum tracking
Description
: ON activates the MAX monitoring function. The maximum value of the channel (not
sensor) is monitored until the value is reset by CALC#:MAX:STAT ON or
CALC#:MAX:STAT OFF is sent.
Syntax
:
CALCulate<channel 1 or 2>:MAXimum[:MAGnitude]?
Example
:
OUTPUT @Pwr_mtr;CALC2:MAX?
! Query channel 2 maximum value in dBm
Response
:
Highest power reading since CALC2:MAX:STAT ON was sent.
Description
: This command reports the value maintained as the maximum in the 8650A max monitor.
7
6
5
4
3
2
1
0
Operational
Status
Require
Service (RQS)
Event Status
Message
Available
0
0
0
0
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...