Series 8035XA Peak Power Sensors
2-8
Manual 21568, Rev. F, March 2008
The default display will be presented after attachment of a new sensor. Sample delay offsets
are available for each sensor. The use of a non-zero offset will be indicated by an asterisk
between the last digit and the units as shown below. The delay displayed is the sample delay
before the application of any offset.
DLY
A
56.345*uS
The Sample Delay display line interacts with the Min/Max display line. If Min/Max is turned on
for a line that Sample Delay was tracking, Min/Max will be displayed instead of Sample Delay.
2.3.2
Setting Sample Delay
The sample delay can be set from two locations within the menu structure. One location is at
the default display where sample delay information is displayed. The other is in the menu
structure where the peak sensor setup is configured.
Sample Delay Operation
You will be presented with a display with a decimal point fixed in the display. Normally, four
digits are available to the left of the decimal point and three digits to the right as shown below.
Leading zeros will be suppressed.
XXXX.XXX
u
S
When the units are set to nS, only one digit will be available to the right of the decimal.
The right and left arrow keys move a cursor to select a digit or unit to be changed. The cursor
will stop only at valid digit or unit locations (valid digits are shown by ^ below). The cursor will
stop under valid blank spaces so that large numbers can be entered quickly.
XXXX.XXX
u
S
^^^^ ^^^ ^
Press the up key to increment the digit value by 1, or the down key to decrement the digit value
by 1. If a digit is incremented past 9, a carry is propagated to the next higher digit (odometer
mode). Similarly, if a digit is decremented past 0, a borrow will be made from the next higher
digit. You can decrement the delay to a negative number.
An exception is made when the cursor is moved to the 10ths of nanoseconds range. The up or
down arrows will change the display in .5 nS increments. This is the maximum resolution of the
sensor.
Press the up or down keys while the cursor is under the units display to cycle through the units.
You will be offered a choice of nS, uS, or mS. The units display does not wrap around.
The units selected will be used in all displays for the delay for the sensor to which they are
assigned.
If a peak measurement parameter is changed, e.g. sample delay or trigger level, the power
display will display
NO TRIG
(see Figure 2-9) until a new measurement is made with the new
parameters in effect. If the sensor triggers quickly, the message may not be seen. In the TR2
Содержание 80350A
Страница 4: ......
Страница 10: ...Series 8035XA Peak Power Sensors vi Manual 21568 Rev F March 2008 ...
Страница 13: ...Manual 21568 Rev F March 2008 ...
Страница 16: ...Series 8035XA Peak Power Sensors xii Manual 21568 Rev F March 2008 ...
Страница 17: ...Manual 21568 Rev F March 2008 ...
Страница 18: ...Series 8035XA Peak Power Sensors xiv Manual 21568 Rev F March 2008 ...
Страница 24: ...Series 8035XA Peak Power Sensors 1 6 Manual 21568 Rev F March 2008 ...
Страница 49: ...Figure 3 5 Digital Timing Diagram INT EXT Trig Mode Theory of Operation Manual 21568 Rev F March 2008 ...
Страница 90: ...Series 8035XA Peak Power Sensors 6 22 Manual 21568 Rev F March 2008 ...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......