Operation
Manual 20791, Rev. C, June 2001
2-61
[
GPIB DEVICES]
Menu SY1:
Accesses menu SY1-10 wherein the GPIB address for the 8003 can be set and the
instrument’s private bus can be turned on or off. The [
SOURCE
] softkey of menu SY1-10 accesses
menu SY1-11. The [
SOURCE ADDRESS
] softkey of menu SY1-11 sets the address of the signal
source (default is 20). The [
AUTOSELECT
] softkey automatically interfaces any compatible
sweeper being used in the system with the exception of a Model HP8350A. If an 8350A is being
used the next softkey down, labeled HP 8350A, should be pressed to activate the interface. The
[
PLOTTER
] softkey of menu SY1-10 brings up menu SY1-12. Through this menu the plotter’s
address can be set (default is 05), and the number of pens the specific plotter contains can be
defined for use by the 8003 when graphic plots are made. The [
PRINTER ADDRESS
] softkey of
menu SY1-10 sets the address of the printer being used in the system (default is 01).
[
CAL SWP RAMP]
Menu SY1:
Synchronizes the 8003 to the ramp from the sweeper. This will need to be done if the
trace length on the display is less than the full 8 horizontal divisions.
[
STANDARD]
Menu SY1-A:
Resets the 8003 to normal (default) operation after the V
PROP
F function has
been used.
[
V
PROP
F]
Menu SY1-A:
Synchronizes the internal measurement process to an external ramp signal.
[
SERVICE]
Menu SY1:
Accesses function testing of various elements of the 8003. This will aid in the main-
taining and servicing of the instrument. This softkey causes menu SY1-13 to appear. The regular
data display will be replaced with a display of the currently defined calibrator output adjustment
and the currently defined serial number of the calibrator.
[
SET CAL]
Menu SY1-13:
Adjusts the 0 dBm output of the calibrator of the 8003. When this softkey is
pressed, the softkey menu will change to menu SY1-14 where the calibrator output and the cal-
ibrator serial number can be revised or modified if required. A password can be defined that
must be entered to enable the adjustment of the calibrator output (see DEFINE PASSWORD
next page). This will be for environments in which access to this ability must be restricted. If a
password is defined for this adjustment function, then all the softkey labels except for the
ABORT label will be in dim, dark blue and cause no action if their associated softkeys are
pressed. The instrument will present a prompt to enter the password. If the correct password is
not entered, the instrument will deny access but will allow the entry to be tried again.
If the password is entered correctly or no password has currently been defined, then all of the
softkey menu labels will be in standard menu green.
[
SET CAL OUTPUT]
Menu SY1-14:
Sets the calibrator output. When this softkey is pressed, the presently defined
value of the calibrator adjustment (in percentage of deviation) will flash and a prompt will be
presented asking for a new adjustment percentage to be entered as desired. A number from
50.00 to 150.00 (%) can then be entered.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...