Operation
Manual 20791, Rev. C, June 2001
2-45
[
DEFINE MRKR N]
First Level Menu F5:
Defines a marker at the cursor frequency. This softkey is also used to turn
markers on and off and to clear markers. When this softkey is pressed, the cursor capability will be
turned on if it is off. The entry area of the display will present a prompt to enter the number
(0 through 9) of the marker that it is desired to define at the cursor frequency. The default number
will be the lowest marker number which has not already been defined. If all markers are currently
defined, then there will be no default number. The softkey menu will change to menu F5-2.
When a number is entered, a marker small circle will be superimposed on the cursor on each trace
and will remain at that frequency when the cursor is moved. If the number of a marker that is
already defined is entered, then the previous frequency definition will be lost and the marker will
automatically be turned on if it was off.
[
DEFINE MRKR N]
Second Level Menu F5-2:
If the DEFINE MRKR N softkey is pressed again, the entry area of
the display will present a prompt to enter the number (0 through 9) of the marker it is desired
to define at the cursor frequency. This softkey will work identically to the DEFINE MRKR N
softkey that led to this sub-menu except that this softkey will not change the softkey menu.
This softkey option is provided in this menu so that one of the other softkeys in the menu
(such as changing the cursor frequency) can be used while still retaining the ability to define a
marker.
[
CURSOR FREQ]
Menu F5-2:
Moves the cursor. Pressing this softkey causes the entry area of the display to indi-
cate both the cursor frequency and the level at the cursor for the entry channel. A prompt will
be presented to move the cursor by changing the frequency with either the spin knob or the
enter keypad, as desired.
This is the same type of data entry that is allowed when the Cursor key is pressed as well as
when the CURSOR FREQ softkey is pressed in the prior menu. It is provided here to allow
both the changing of the cursor and the defining of markers from one softkey menu.
[
MRKR N ON] and [MRKR N OFF]
Menu F5-2:
Turns any defined markers off and on without having to redefine their frequen-
cies. When a marker is turned off, its tic mark is no longer displayed on the measurement traces
and the cursor can no longer be assigned to the marker. This allows markers to be set up for dif-
ferent tests while only displaying the relevant markers for each particular test, as desired.
When one of these softkeys is pressed, the entry area of the display will present a prompt to
enter the number of the marker that it is desired to turn on or off. The default will be to turn all
defined markers on or off. This would occur if the enter key were pressed without entering a
number. Note that only defined markers can be turned off or on. An error message will be gen-
erated if the number of an undefined marker is entered.
If the sweep frequencies were changed so that the frequency of a defined marker is outside the
limits of the frequency range currently being swept, then the marker would automatically be
turned off. Such a marker cannot be turned back on again until the sweep frequencies are
changed to allow the marker to be displayed.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...