Performance Verification
Giga-tronics 2500B Series Microwave Signal Generators
Page 118
Operation Manual, Part Number 34737, Rev C, June 30, 2011
9.2.11
Frequency Modulation Accuracy - Low Rate
This procedure tests the frequency modulation accuracy of the 2500B in low rate.
9.2.11.1
Equipment and Materials
Microwave spectrum analyzer
Function generator
Digital voltmeter
Cables and adapters for connecting the test setup
Datasheet 8 (Table 76 on page 120)
Figure 53: Frequency Modulation Accuracy and Deviation Bandwidth Test Setup
NOTE:
Set the output amplitude (into 50 Ohms) of the function generator accurately using a precision digital voltmeter.
NOTE:
The external frequency reference (timebase) should be “Tee” connected from the 2500B to both the spectrum
analyzer and the function generator.