Performance Verification
Giga-tronics 2500B Series Microwave Signal Generators
Page 110
Operation Manual, Part Number 34737, Rev C, June 30, 2011
9.2.8
Step Attenuator Level Accuracy
This procedure tests the accuracy of the RF power output of the 2500B when the step attenuator option is included.
NOTE:
When measuring the flatness and accuracy of the 2500B, consider the various measurement uncertainties in the
test system. These include, but are not limited to, VSWR, cable loss, and linearity.
NOTE:
Verify that the amplitude accuracy specification for the specific spectrum analyzer used has adequate
performance capability for this test.
9.2.8.1
Equipment and Materials
Microwave spectrum analyzer
Cables and adapters for connecting the test setup
Datasheet 5 (Table 70 on page 112)
Figure 51: Step Attenuator Level Accuracy Test Setup