Weighing parameters setting menu
Autozero at power up
This parameter is the maximum weight zeroed at power up. The auto-zero func-
tion consists in carrying out an automatic zero calibration when the instrument is
switched on, only if the detected weight stabilizes within the set threshold. To
disable the function, set the value 0.
Tracking of zero
When the scales are discharged, the instrument automatically corrects small and
slow weight variations (positive or negative) compared to the zero scale.
Value
Variation
0
Tracking function of zero excluded.
1
Minimum zero tracking intervention
2
Medium zero tracking intervention. (standard)
3
High zero tracking intervention.
4
Maximum zero tracking intervention.
Serial menu
In this menu it is possible to select the baud rate and the format of the continuous serial transmission.
Select baud rate
Select the baud rate to be used for continuous transmission on COM1 in RS232
or RS485 (selections: 2400, 9600, 19200, 38400, 57600, 115200)
Select data format
Select the data format to be used for continuous transmission on COM1 in RS232
or RS485 (selections: N-8-1, O-8-1, O-7-1, E-8-1, E-7-1, N -8-2)
0 trAC
Auto 0
bAud
ForMAt
Pagina –
page
–
Seite
17