GHM GREISINGER GMK 100 Скачать руководство пользователя страница 7

H68.0.01.6C-06 

 Operating Manual GMK 100 

page 7 of 14 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

 

6.3  Hold the device 

The water content of the hand has an impact on the measuring value, if the device is held 
inappropriately.   
The best results will be obtained if you lay the device down or hold it the way as shown in picture 3.  

 

 

 
 
 
 

 

 Picture 1: Held in a wrong way! 

Picture 2: laid down 

– right! 

Picture 3: Held in the right way!    

6.4  Characteristics and materials 

Material 

Recommended depth switch setting 

rEF 

 

Reference characteristics 

 

d.45 

Wood, density = 450 kg/m

3

 

 

d.50 

Wood, density = 500 kg/m

3

 

 

d.55 

Wood, density = 550 kg/m

3

 

 

d.60 

Wood, density = 600 kg/m

3

 

 

d.65 

Wood, density = 650 kg/m

3

 

 

d.70 

Wood, density = 700 kg/m

3

 

 

d.75 

Wood, density = 750 kg/m

3

 

 

d.80 

Wood, density = 800 kg/m

3

 

 

d.85 

Wood, density = 850 kg/m

3

 

 

d.90 

Wood, density = 900 kg/m

3

 

 

d.95 

Wood, density = 950 kg/m

3

 

 

d.99 

Wood, density = 1000 kg/m

3

 

 

bEt 

Concrete (density 

≈ ca. 2.2 to/m

3

25 mm 

E.ZE 

Cement screed (density 

≈ 2.2 to/m

3

25 mm 

E.ZC 

Cement-calcium sulfate-flowing screed  25 mm 

E.An 

Anhydrite screed 

25 mm 

P.Gi 

Gypsum plaster 

10 mm 

P.2 

Lime cement plaster 

10 mm 

You can find a list of different types of wood and their corresponding characteristics d.xx in the 
addendum.  
The materials are selected with these keys: 

 material selection upwards: press 

  

 

 Material selection downwards: press 

 at the same time 

 

The use of inappropriate characteristics can cause faulty measurements! 

 

 

Содержание GREISINGER GMK 100

Страница 1: ...402 9383 0 info greisinger de www greisinger de Operating Manual Capacitive moisture measurement device for non destructive measurement as of version 1 3 GMK 100 Please carefully read these instructions before use WEEE Reg Nr DE 93889386 Please consider the safety instructions Please keep for future reference ...

Страница 2: ...REAR SIDE 5 5 START OF OPERATION 5 6 MEASUREMENT BASICS 6 6 1 MEASURING FIELD DEPTH 6 6 2 MOISTURE RATING WET MEDIUM DRY 6 6 3 HOLD THE DEVICE 7 6 4 CHARACTERISTICS AND MATERIALS 7 6 5 ZEROING FUNCTION 8 7 WOOD MEASURING 8 8 PLASTER MEASURING 8 9 FLOOR SCREED MEASUREMENT 8 10 PLASTERBOARD MEASUREMENT 9 11 MEASUREMENT OF OTHER MATERIALS 9 12 ADDITIONAL INFORMATION TO MOISTURE MEASUREMENT 9 12 1 MOI...

Страница 3: ...elp for this The instructions of the manual have to be understood regarded and followed To be sure that there s no risk arising due to misinterpretation of measured values the operator must have further knowledge in case of doubt the user is liable for any harm damage resulting from misinterpretation due to insufficient knowledge The manufacturer will assume no liability or warranty in case of usa...

Страница 4: ...ill be adhered to when using the device 1 Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification If the device is transported from a cold to a warm environment condensation may cause in a failure of the function In such a case make sure the device temperature has adjusted to t...

Страница 5: ...r 2sec Zeroing function p r t chapter 6 5 Key 2 and 3 together sort down Material selection downwards p r t chapter 6 4 Depth switch on the right side Sliding switch up ca 10 mm measuring depth d 10 Sliding switch down ca 25 mm measuring depth d 25 After changing the measuring depth d 10 for 10mm or d 25 for 25 mm is briefly displayed 4 3 Measuring spot rear side Rear side of device The entire lig...

Страница 6: ...ar side the area is marked under that it is predominantly measured For precise measurements the device has to lie on the measured material not only at the marked measuring spot but at its whole rear side If there is an air gap between device and material the measured value will be too dry Here are some examples for measurements that are not precise at all Measured value too dry in all cases False ...

Страница 7: ...3 d 50 Wood density 500 kg m3 d 55 Wood density 550 kg m3 d 60 Wood density 600 kg m3 d 65 Wood density 650 kg m3 d 70 Wood density 700 kg m3 d 75 Wood density 750 kg m3 d 80 Wood density 800 kg m3 d 85 Wood density 850 kg m3 d 90 Wood density 900 kg m3 d 95 Wood density 950 kg m3 d 99 Wood density 1000 kg m3 bEt Concrete density ca 2 2 to m3 25 mm E ZE Cement screed density 2 2 to m3 25 mm E ZC C...

Страница 8: ...kind of wood is not listed in addendum A you have to know its density dry for example a wood of density 0 68 kg dm3 is measured with material d 70 approximate actual density to the next material 0 52 d 50 0 53 d 55 8 Plaster Measuring Rough surfaces will influence the measuring results Precise values are only obtained at even surface and without any air gap between surface and device Metal grids c...

Страница 9: ...ring values 12 Additional Information to Moisture Measurement 12 1 Moisture u and water content w Either moisture u or water content w is needed according to the application Carpenter joiner etc normally use moisture u relating to oven dry mass Water content w is normally used for combustibles i e wood briquette You can choose whether the device should use moisture or water content p r t chapter 1...

Страница 10: ...to the next parameter by pressing Parameter Value Information button buttons P OF Auto Power Off turn off delay factory setting 20 min 1 120 Auto Power Off turn off delay in minutes If no key is pressed for the time adjusted here the device is automatically switched off adjustable 1 120 min OF Auto power off is deactivated continuous operation UNI Display unit factory setting u u Measuring value d...

Страница 11: ...alue rEF 1 slope correction 100 Follow these instructions to adjust the device Switch the device off Press and keep it pressed while turning the device on press shortly Release the hold button not before the first parameter S 10 is displayed Set parameter with up or down Jump to the next parameter by pressing Parameter Value Information button buttons Slope correction for measuring depth 10 mm fac...

Страница 12: ...emperature The blinking bAt display indicates low battery voltage device will continue to work for a short time The battery is consumed and has to be changed Measurements are no longer possible 17 Reshipment and Disposal 17 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or sensor...

Страница 13: ...l Display 2 displays for material and measured value backlight Hold function Press button to freeze current value Working temperature 25 to 50 C 0 to 80 RH non condensing Storage temperature 25 to 70 C Power supply 9 V battery type IEC 6F22 included Power consumption measurement Approx 0 12 mA battery life time more than 2500 hours for alkaline battery Power consumption backlight Approx 2 5 mA Bac...

Страница 14: ...acuminata grandiflora d 50 Mahogany Swietenia spp d 50 Mahogany African Khaya spp d 50 Mahogany Philippine Parashorea plicata Shorea almon d 50 Maple New Guinea Flindersia pimentelianan d 55 Maple Sugar Acer saccharum d 70 english lat charact Maple Sycamore Acer pseudoplatanus d 55 Mengkulang Heritiera spp d 65 Meranti Dark Red Shorea spp d 65 Meranti White Shorea hypochra d 55 Meranti Yellow Shor...

Отзывы: