7 | Error and system messages
G 1610
22 / 26
B-H86.0.2X.DK2-3.0
7
Error and system messages
Display
Meaning
Possible
causes
Remedy
----
Sensor cable
defect
Sensor or device
defect
Measuring range
exceeded/undercut
Cable breakage
Sensor or device
defect
Measurement outside
permissible range
Send in for repair
Send in for repair
Stay within allowable
measurement range
No display,
unclear characters
or no response
when buttons are
pressed
Battery depleted
System error
Device is defective
Battery depleted
Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too high
Faulty calibration
Electrode or device
defect
Stay within allowable
measurement range
Perform calibration
Check electrode
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too low
Electrode or device
defect
Check electrode
Send in for repair
SYS ERR
System error
Error in the device
Switch device on/off
Replace batteries
Send in for repair
(AL ERR.2
Slope is too low
Incorrect oxygen
reference
Electrode contaminat-
ed or defective
Perform calibration in
damp environmental air
Maintain the electrode
(AL ERR.3
Slope is too high
Incorrect oxygen
reference
Electrode contaminat-
ed or defective
Perform calibration in
damp environmental air
Maintain the electrode