G 1409
Operation | 7
B-H87.0.01.DB2-2.1
21 / 28
Call up menu
Hold
Release
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
7.2.5
Configuring parameters of the adjustment menu
The following representation shows the available parameters and various configura-
tion options.
The
Adjustment
menu is open. See
Adjustment of the measuring input [
1.
Select the desired parameter you would like to configure.
2.
Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3.
The available configuration options are listed for each parameter in the following
representation.
Parameter
Values
Meaning
Zero point correction
T.OF
0.00
No zero point correction
-5.00 .. 5.00
Zero point correction in °C. and/or at °F -9.00 ..
9.00
Gradient correction of the temperature
T.SL
0.00
No gradient correction
-5.00 .. 5.00
Gradient correction in %
Gradient correction for the conductivity value
T.SL
1.000
No gradient correction
0.800 .. 1.200
Multiplier for the gradient correction
Zero point correction:
Displayed value = measured value
–
T.OF
Representation
Outcome of an action
Description
Prerequisites
Instruction
Representation
Formula