G 1400 series
Error and system messages | 7
B-H87.0.0X.DK2-2.1
13 / 20
7
Error and system messages
Display
Meaning
Possible
causes
Remedy
----
Range switches or
value unstable
Measurement far
outside of range
Controlling instable
Measuring cell defect
Contamination or air
bubbles
Wait for the transient effect of
the controller
clean contamination /bubbles
Keep measurement within the
permissible range
Send in for repair
(, , )
SENS ERRO
Cable defect
Sensor or probe
defect
Measuring range
exceeded or undercut
Cable breakage
sensor or probe defect
Measurement outside
range
Send in for repair
Send in for repair
No display,
unclear charac-
ters or no
response to
buttons
Battery depleted
System error
Product defect
Battery depleted
Error in the product
Product defect
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too high
Measuring cell defect
Stay within allowable meas-
urement range
Check the measuring cell/
send in for repair
ERR.2
Measuring range is
undercut
Measurement too low
Measuring cell defect
Stay within allowable meas-
urement range
Check the measuring cell/
send in for repair
SYS ERR
System error
Error in the product
Switch product on/off
Replace batteries
Send in for repair